Page 9
WASTE CONTAINER MODULES
Saliva and High Volume Waste Container
FLOAT SENSOR CONNECTOR (PN 860080):
Disassemble Float Sensor Connectors by trimming
back heat shrink and cutting wires to Connectors (see
Fig. 23a). Remove nuts that hold Connectors to
Plumbing Chassis.
To reassemble, place Float Sensor Connector through
Plumbing Chassis. Apply RTV to Connector to avoid
leakage. Slide sleeving over wires. Attach nut to Sensor.
WASTE CONTAINER FLOAT SENSOR (PN
730514):
Remove nut (see Fig 23b). Cut wires and remove
Sensor.
To reassemble, resolder new Sensor wire and apply
heat shrink. Reattach nut.
PLUMBING CHASSIS - Cont’d
SYRINGE (PN TA-90D):
Unscrew handle from Syringe head (see Fig. 22).
Remove tube clamps and tubes from fittings on Syringe
head. Remove tube clamps from tubes and remove
handle.
Reassemble by placing tube through handle and plac-
ing tube clamps on tube. Place water tube (designated
by ribbed lines on side of tube) to the water side. Water
side is indicated by water symbol on head. Attach han-
dle to Syringe head and test for air, water, and spray.
Figure 21
Blue
Wire
Sensor Bracket
Relay
Tube to
Pressure
Sensor
Diodes
Sensor
Mtg Nut
Pressure
Sensor
Sensor
Tube Fitting
Handle
Head
Figure 22
Figure 23a
Float Sensor
Connectors
Figure 23b
Nut
Container
Lid
Float
Sensor
Float Sensor
Connector
Water
Symbol
Baffle