2.
SETTING UP THE UNIT:
1. Unpack the console and check that the unit
is compatible with the voltage from the
power source. Use the 110V unit for 110-
120V 60Hz voltages, and the 220V unit for
220-250V 50Hz voltages. For alternative
voltages (220V-250V), confirm that the
type of plug cap is correct for the country
of usage and carries the proper certifica-
tion markings. The correct cord is shipped
with the console.
2. Attach the remote power cord to the back
of the console and plug into a grounded
electrical receptacle.
3. Connect the M-8 10K motor to the recepta-
cle on the lower front panel of the console.
4. Attach the appropriate “E” Type handpiece
to the motor, i.e. a 1:8 reduction contra
angle.
5. Attach the supplied foot control to the con-
nector on the back of the unit marked
“Foot Switch”. An On/Off type or a variable
speed type foot control may be used.
MOTOR
NOTE: The hand motor will remain sufficiently cool
during the normal intermittent use (IEC 601 defined
frequency of use) for which it was designed.
“E” TYPE
HANDPIECE
POWER
CORD
HANDPIECE
CONSOLE
M8-10K
MOTOR
FOOT SWITCH