Water On/Off Toggle:

Controls the on/off flow of water coolant to

the handpiece.

Water Flow Control:

Adjusts the water coolant flow to the

handpiece.

Pressure Gauge:

Gives a visual indication of the drive air

pressure to the handpiece.

Drive Air Pressure Control:

Adjusts the drive air pressure to the

handpiece.

Foot Control:

Controls the drive air pressure to the

Figure 3

DEPRESS

LOCKING COLLAR

handpiece.  Apply light foot pressure to

any part of the foot control cover.

3-Way Air /Water Syringe:

Delivers air, water, or a combination mist.

Supplied with autoclavable tip.

NWS-7 Water System (Optional)

:

White 1-liter bottle dispenses water

through the  3-Way Air/Water Syringe and

handpiece.

AA-61 Water Connector

:

Provides water via standard 1/4” quick

disconnect to independent Cavitron or

scaler units requiring a water supply.

OPERATION FUNCTIONS: 

(Refer to Fig. 2)

UNIT OPERATION:

Handpiece

- Connect either a high-speed or

low-speed handpiece to the handpiece

connector. Depress foot control for operation.

3-Way Air/Water Syringe

- Depress the right

button for air operation, and the left button for

water operation. Depressing both buttons will

create a mist. The syringe features quick-

change autoclavable tips: To remove a tip,

press on the locking collar surrounding the tip

socket and pull the used tip straight out of the

socket (see Fig 3). To insert a new tip, press

locking collar and push tip into socket as far

as it will go. Release collar and gently tug on

tip before using to ensure that tip is securely

locked into socket.

CLEANING

Syringe Tip Sterilization:

1)

Remove contaminated syringe tip.

2)

Remove all visible signs of contamination

before autoclaving.

3)

Autoclave tip at 132° C (270° F) for ten minutes.

4)

Sterilize between each patient use.

NOTE

: Since only the tips can be autoclaved,

it is recommended that the air/water syringe

be bagged with a disposable, single-use

plastic sleeve between each patient use.

Water Lines:

Disinfect the water lines weekly.

Prepare a 1:10 bleach solution (1 part

household bleach to 10 parts water). Remove

water reservoir and discard residual water.

Replace empty water supply tank and air

purge all waterlines. Fill water supply tank with

bleach solution. Run bleach solution through

all lines. Allow bleach solution to stand in lines

for 10 minutes. Remove water supply tank and

discard bleach. Flush water supply tank and

all lines thoroughly with clean water. Air purge

and leave lines dry until next clinical use.