Water On/Off Toggle:
Controls the on/off flow of water coolant to
the handpiece.
Water Flow Control:
Adjusts the water coolant flow to the
handpiece.
Pressure Gauge:
Gives a visual indication of the drive air
pressure to the handpiece.
Drive Air Pressure Control:
Adjusts the drive air pressure to the
handpiece.
Foot Control:
Controls the drive air pressure to the
Figure 3
DEPRESS
LOCKING COLLAR
handpiece. Apply light foot pressure to
any part of the foot control cover.
3-Way Air /Water Syringe:
Delivers air, water, or a combination mist.
Supplied with autoclavable tip.
NWS-7 Water System (Optional)
:
White 1-liter bottle dispenses water
through the 3-Way Air/Water Syringe and
handpiece.
AA-61 Water Connector
:
Provides water via standard 1/4” quick
disconnect to independent Cavitron or
scaler units requiring a water supply.
OPERATION FUNCTIONS:
(Refer to Fig. 2)
UNIT OPERATION:
Handpiece
- Connect either a high-speed or
low-speed handpiece to the handpiece
connector. Depress foot control for operation.
3-Way Air/Water Syringe
- Depress the right
button for air operation, and the left button for
water operation. Depressing both buttons will
create a mist. The syringe features quick-
change autoclavable tips: To remove a tip,
press on the locking collar surrounding the tip
socket and pull the used tip straight out of the
socket (see Fig 3). To insert a new tip, press
locking collar and push tip into socket as far
as it will go. Release collar and gently tug on
tip before using to ensure that tip is securely
locked into socket.
CLEANING
Syringe Tip Sterilization:
1)
Remove contaminated syringe tip.
2)
Remove all visible signs of contamination
before autoclaving.
3)
Autoclave tip at 132° C (270° F) for ten minutes.
4)
Sterilize between each patient use.
NOTE
: Since only the tips can be autoclaved,
it is recommended that the air/water syringe
be bagged with a disposable, single-use
plastic sleeve between each patient use.
Water Lines:
Disinfect the water lines weekly.
Prepare a 1:10 bleach solution (1 part
household bleach to 10 parts water). Remove
water reservoir and discard residual water.
Replace empty water supply tank and air
purge all waterlines. Fill water supply tank with
bleach solution. Run bleach solution through
all lines. Allow bleach solution to stand in lines
for 10 minutes. Remove water supply tank and
discard bleach. Flush water supply tank and
all lines thoroughly with clean water. Air purge
and leave lines dry until next clinical use.