Aseptico AA-98 AirPac II Скачать руководство пользователя страница 4

3

SET-UP & OPERATION

1. Unpack the compressor and locate the

remote power cord.

2. Plug the power cord into the receptacle on

the outside of the unit and then into a
grounded power outlet.

3. To activate the AA-98 compressor, flip the

power switch to the on (I) position.  Once
powered, the compressor will start and stop
automatically to maintain pressure.  The
pressure switch is factory adjusted to
properly start and stop.  The cooling fan will
operate continuously.

4. After plugging the compressor in and turning

it on, let the compressor run until it stops,
indicating full pressure.

5. Locate the 1/4" quick disconnect air outlet

and pressure gauge on the end of the
compressor case.  Connect a 1/4" male
coupling from your equipment to the air
output fitting.  The pressure gauge indicates
pressure in the air storage tanks and gives a
visual indication that the compressor is
maintaining pressure properly.

6. The AA-98 compressor will cycle on and off

as air is consumed during use.  This is
normal for operation.  Turn compressor off if
it is not to be used for extended periods of
time.

7. To turn the compressor off, flip the power

switch to the off (O) position.

AIR

OUTLET

PRESSURE

GAUGE

10A CIRCUIT

BREAKERS

POWER CORD

RECEPTACLE

POWER

SWITCH

Содержание AA-98 AirPac II

Страница 1: ...AA 98 AirPac II Compressor 220V OPERATION and MAINTENANCE INSTRUCTION MANUAL ...

Страница 2: ...eat even under normal operating conditions To avoid serious burns never touch the compressor during or immediately after operation 4 When unit is not in use store the cord wrapped around the carry handle Do not abuse cord 5 Servicing should always be performed by an Authorized Service Center 6 ALWAYS use a grounded 3 prong plug in P O Box 1548 Woodinville WA 98072 425 487 3157 800 426 5913 www ase...

Страница 3: ...and an automatic pressure switch control Congratulations This system is engineered to provide many years of reliable service Please read the instructions provided in this manual to receive the best and longest service from your Aseptico equipment Separate manuals may be provided to cover the operation and maintenance of handpieces or other accessories for your unit PACKAGE CONTENTS Compressor Carr...

Страница 4: ...t the compressor run until it stops indicating full pressure 5 Locate the 1 4 quick disconnect air outlet and pressure gauge on the end of the compressor case Connect a 1 4 male coupling from your equipment to the air output fitting The pressure gauge indicates pressure in the air storage tanks and gives a visual indication that the compressor is maintaining pressure properly 6 The AA 98 compresso...

Страница 5: ...erside of the compressor Carefully and quickly turn open the drain fitting Return the unit to its normal position and allow water to drain out with the escaping air AIR FILTER A dirty air filter will not allow the air compressor to operate at full efficiency If a decrease in the supply air is noticed it may be necessary to replace the filter element The air filter is located inside the case along ...

Страница 6: ...e a low voltage situation Low voltage can be suspected when 1 The motor does not achieve full power or speed 2 Fuses or circuit breakers activate when starting the compressor 3 Lights dim when the compressor is started 4 Other electrical appliances fail to operate properly 5 Compressor shakes or hesitates when starting A primary cause of voltage drop is using an extension cord that is not heavy en...

Страница 7: ...cylinder 3 4 hp Oilless compressor 100 PSI Displacement 4 1 CFM 116 LM Electrical 220 volt AC 50 Hz Air Delivery 1 20 CFM at 100 PSI 60 LM at 689 KPa 1 85 CFM at 50 PSI 74 2 LM at 345 KPa Circuit Breakers 10 Amps Quantity 2 Amps at Working Pressure 5 1 Amps Maximum pressure 100 PSI Air Storage Capacity 1 0 Gal Weight 43 lbs Size 21 5 x 8 5 x 12 75 ...

Страница 8: ...ent or product in part or whole Aseptico shall be the sole arbiter of such action In the event of alleged defect under warranty the purchaser is to notify Aseptico s Customer Service Department promptly Customer Service will provide instructions usually directing that the product be returned for service Shipment to Aseptico and the cost thereof is always the responsibility of the purchaser Acciden...

Отзывы: