DFU-0332r0_fmt_en-US
Page 12 of 64
Figure 2-
Synergy
ID
Rear Panel
1.9.2
Rear Panel
1.
DVI Video Output Connector
-Supplies a
digital video signal output in DVI-D format.
2.
Laser Input-
Connection for Synergy Laser
Light Source. Also acts as second DVI output
if laser is not connected.
3.
Display Ports (2X)-
Supplies UHD Video
Signal output in either 1.1 [dual cable] or 1.2
[MST].
NOTE: Arthrex recommends connecting
Synergy
ID
to the primary surgical monitor via
multiple output types (e.g., display port and
DVI, Synergy Matrix and 3G-SDI) in the event
that one type of connection is lost.
4.
Audio In / Audio Out-
Audio In: Line Level
Audio input for Microphone. Audio Out: Line
Level Audio output to Medical Grade Devices.
5.
RS-232 Connector
-Isolated connection to
devices requiring Serial Control.
6.
DVI Input-
1080P/60 Input for Picture in Picture
[PIP].
7.
USB 2.0 (2X)-
USB 2.0 Connection.
8.
USB 3.0 (2X)-
USB 3.0 Connection.
9.
Accessory Ports (Inputs/Outputs (2X) mini
Stereo-Phone Connectors)
-Accessory ports
allow for control of the CCU with a footswitch
or external device or for the CCU to control
external devices via the camera head buttons.
10.
Tablet Connection-
Connection for Tablet
Data Input device. Provides for data
interchange and tablet charging.
11.
3G-SDI Out
-1080P/60 Output.
12.
Ethernet Connector
-Isolated 10/100 Mb/sec.
13.
Potential Equalization Connector (POAG)
-
Potential Equalization Connector per DIN
42801.
NOTE: The purpose of the Potential
Equalization Connector is to equalize the
potentials between different metal parts of the
various Medical Electrical [ME] equipment
which make up a Medical Electrical system, or
to reduce differences of potential which can
occur during operation between the bodies of
the Medical Electrical devices and conductive
parts of other objects. The Potential
Equalization Connector may be connected
directly between any ME Devices, or to a
common busbar of the electrical installation.
Reference IEC 60601-1 for ME Systems.
14.
IEC 320 Power Inlet Module (100-240V~,
50/60 Hz)
-The CCU is equipped with a
switching power supply that automatically
adjusts to the line voltage being used. Accepts
the supplied hospital grade power cord.