200 Micron Blood Component Filter and Syringe Adapter
Instructions for use
Prime heparin flushed filter and syringes. Residual priming volume
of saline is 6 mL. Maximum volume of product is 300 mL.
Directions:
Use aseptic technique. Single patient use.
1.
Close the pinch clamp on the distal end of filter tubing.
2.
While holding adaptor upright, squeeze the filter chamber
and attach the syringe containing the blood component to the
proximal end.
3. Release hold on chamber allowing blood components to enter
filter chamber. Release pinch clamp and allow blood compo
-
nents to flow into distal tubing just to luer connector to de-air
tubing and close pinch clamp.
4. Attach the collection syringe to distal tubing luer connector of
the filter and open pinch clamp.
5. Gently pull back on the plunger of the collection syringe until
resistance is met, or until the desired volume is reached. If re-
sistance is met before desired volume is reached, disconnect
proximal syringe and pull remaining blood components into
collection syringe. Repeat if more than one syringe of blood
component is drawn.
6.
Disconnect the filtered blood component syringe from filter tub
-
ing and attach a sterile syringe cap.
Follow AABB guidelines.
Manufacturer:
Arteriocyte Medical Systems, Inc.
45 South Street
Hopkinton, MA 01748
USA
Internet: www.arteriocyte.com
Toll-free: 1-866-660-AMSI (2674)
Fax: 508-497-8951
Содержание magellan
Страница 4: ...LQFKHV PP PP 0 AMS Confidential AMS95030LBL Rev 2 AMS95125LBL Rev 2...
Страница 6: ...LQFKHV PP PP 2 AMS Confidential 0 AMS95030LBL Rev 2 AMS95125LBL Rev 2...
Страница 13: ...LQFKHV PP PP AMS Confidential AMS95030LBL Rev 2 AMS95125LBL Rev 2...
Страница 15: ...MAGELLAN Aspiration Kit Instructions For Use...
Страница 16: ......
Страница 25: ......