![Array electronic 367 Series Скачать руководство пользователя страница 66](http://html1.mh-extra.com/html/array-electronic/367-series/367-series_user-manual_2975638066.webp)
66
Voltage Programming Speed
Maximum time required for output voltage to settle within 1% of its total excursion
(for resistive load). Command processing time is excluded.
3672A 3673A 3674A 3675A
Full load Up
50msec 50msec 60msec 60msec
Full load Down
50msec 50msec 60msec 60msec
No load Up
50msec 50msec 60msec 60msec
No load Down
200msec 300msec 300msec 500msec
AC Input Ratings
AC180V-265V 47Hz
~
63Hz 1050VA
Max
Operating Temperature
0
~
40
℃
0
~
80%RH
Cooling
Fan Cooling
Output Voltage Overshoot
<1V
Programming Language
SCPI (Standard Commands for Programmable Instruments)
Recommended Calibration Interval
1 year
Net Weight
5.5 kg