4 Preparing the device for operation
26
Instructions for Use, AquaVENT
®
FD140, Rev. 12
Warning
Ensure that only dry and clean gases are used. Water, oil
or particles entering the device can damage the device
and endanger the patient.
4.3.2 Single-gas operation
The design of the AquaVENT
®
FD140 allows it to be operated with a single
gas supply at any time. To avoid alarm messages, the following must be
observed:
If an AIR gas supply is not available, the FiO
2
must be set at 100%, if an O
2
gas supply is not available, at 21%.
Please refer to chapter 6.1 Alarms for details of the alarm system.
4.4 CPAP breathing circuit
The AquaVENT
®
FD140 FD140 is fully compatible with all CPAP breathing
circuits and components specified in chapter 10.2 Accessories. Please refer
to chapter 1.1 About this manual for details.
Caution
While gas flows through the respiratory gas outlet, do not
obstruct the outlet to prevent pressure build-up inside the
device.
Содержание AquaVENT FD140
Страница 1: ...AquaVENT FD140 Instructions for use CRITICAL CARE...
Страница 6: ...0 Contents 4 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Страница 11: ...Instructions for Use AquaVENT FD140 Rev 12 9 2Safety In this chapter 2 1 Safety precautions 10...
Страница 16: ...2 Safety 14 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Страница 56: ...5 Operation 54 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Страница 66: ...6 Alarms and messages 64 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Страница 70: ...7 Cleaning 68 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Страница 76: ...8 Maintenance 74 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Страница 88: ...9 Technical specifications 86 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Страница 95: ...Instructions for Use AquaVENT FD140 Rev 12 93 This page is intentionally blank...