MAINTENANCE
ARMEDICA™ Treatment Tables are equipped with a maintenance free electric motor. The moving parts
of the table should have a drop of oil placed on them approximately every six months.
Frequently check to make certain that all hardware (nuts, bolts, etc.) are properly adjusted and securely
fastened.
In the event it becomes necessary to replace the switch cord or the power cord, use the gray cord
removal tool included with this manual to remove cord. Make certain that the plug is in properly and the
locking tabs on the motor are snapped into place.
The vinyl upholstery should be cleaned with mild soap and water and the table frame wiped with a dry
cloth to remove dust and lint when necessary.
SYMBOLS AND INDICATORS
Attention: Consult accompanying documents.
Type B Equipment: “An adequate degree of protection against electric shock is provided,
particularly regarding leakage currents and reliability of the protective earth connection.”
The grounding pin located on the power supply cable is for functional use only, and is
present for EMI/ EMC purposes. The product is Class II equipment and does not rely on
this grounding means for safety.
TROUBLE SHOOTING
MOTOR DOESN’T WORK.
A.
Check to see if motor is plugged into receptacle, and the power cord is properly plugged into the
motor.
B.
Make sure receptacle is getting power.
C.
Check switch cable running from switch box and make sure it is plugged in properly.
D.
Thermal cutout may have activated. Allow the motor to cool.
E. In the event there is a failure of the motor, the motor must be returned to Armedica Mfg. Corp.
The motor must be serviced by the manufacturer.
MOTOR SPECIFICATIONS
TRANSPORT AND STORAGE
Magnetic Corp., Olney, Illinois 62450
Ambient temperature: -20 to 70deg. C
Model MAX 65-A200415A1001-000.
Relative humidity: 10 to 100%, including
Rated Voltage and Frequency 120V 60HZ.
Condensation
Current 1.8 Amps.
Atmospheric pressure: 500hPa to 1060hPa
UL CORD SET 140-355
Duty cycle: on/off Int. 1 min. / 9 min.
Micro switch model E13-00K
TABLE LIFTING CAPACITY: 500 LBS. (226 KILOGRAMS)
2