AC V / DC V
Caution
When connecting the test leads to the
circuit or device, connect the black
lead before the red lead ; when
removing the test leads, remove the
red lead before the black lead.
Resistance
Continuity
Diode
DCA Zero
ACA / DCA
Remove the Jaw out of the conductor.
Press KEY > 1sec to compensate the
residual magnetism.
Caution
Remove Jaw out of the conductor before
Powering up. The Meter will execute DCA
Zero during powering up.
Priority of function
The LCD display shows
voltage mode when the
input Terminal is satisfied
by the following conditions:
ACV 1.3V ~ 999.9V
DCV 2.1V ~ 999.9V
-0.7V ~ -999.9V
The LCD display shows
Ω
mode when the input
terminal is satisfied by the
following conditions:
0 ~
∞
Ω
ACV 0 ~ 0.9V
DCV -0.4 ~ -0.2V
DCV 1.0
~
2.0V
The LCD display shows
diode mode when the
Input Terminal is satisfied
by the following conditions:
DCV 0.4 ~ 0.8V
The LCD display shows
AC mode when the
Current Input Jaw is
satisfied by the following
conditions :
ACA 0.9A ~ 600.0A
DCA 0.9A ~ 600.0A
CAT.
Ⅳ
.600V
with respect to
earth for the jaw.
Barrier for
Hand Guard
Do not hold
the meter above
the Barrier
•
Input Impedance
≧
4K for input voltage up to 30V.
Impedance increase with input voltage
to approximate 375K
Ω
at 750V.
Warning
Do not apply more than 1000VDC / 750
VAC between Tester terminal and earth
ground.