Message
What it means
What to do
A used test strip was inserted.
Voice prompt:
“Used test
strip. Please repeat the test
with a new test strip.”
Repeat the test with a
new test strip.
The blood or control solution
sample was applied before the
symbol appeared.
Voice prompt:
“Error 2.
The sample has been applied
before the blood symbol
appeared. Please repeat the
test with a new test strip.”
Repeat the test with
a new test strip and
wait until the
symbol appears before
applying the blood or
control solution sample.
The temperature during the
test was above or below the
operating range.
Voice prompt:
“Error 3.
The temperature is out of
the operating range. Please
move to an area where the
temperature is within the
operating range and repeat
the test after the meter has
reached a temperature within
the operating range.”
Move to an area where
the temperature is
within the operating
range 42.8–111.2ºF
(6–44ºC) and repeat the
test after the meter and
test strips have reached
a temperature within the
operating range.
The blood sample has
abnormally high viscosity or
insufficient volume.
Voice prompt:
“Error 4.
The sample was not enough
or too thick. Please repeat
the test with a new strip.”
Repeat the test after
inserting a new test
strip.