54
www.arkrayusa.com
7
Repeat the same procedure for the
lancing device (step
1
to step
6
).
NOTE
• After the pre-cleaning and disinfection procedure, the control
solution should be tested to confirm that the meter works
properly before using the meter.
• Control solution tests should be performed with two different
levels of GLUCOCARD Shine Control Solutions (Level 1 and 2).
• Verify that the test results are within the range printed on the
test strip vial.
• See pages 29-31 for how to do a control solution test.
NOTE
If any of the following deterioration signs appear
after pre-cleaning or disinfecting, please stop using the
system and contact ARKRAY Technical Customer Service:
800.566.8558, 24 hours a day, 7 days a week.
• The inscriptions on the exterior of the meter or lancing
device have been removed.
• The color of the meter or lancing device has changed
or faded.
• Cracks or roughness develop on the meter or lancing
device.
• A part of the segment on the meter display does not flash.
• Control solution test results do not fall within the stated
range on the test strip vial.