66
5. To disinfect the GLUCOCARD Expression meter, dispense a second
towel (DISPATCH) from its package. Wipe surface with towel
(DISPATCH) until completely wet. You must wipe all external areas of
the meter including both front and back surfaces and allow 1 minute
contact at room temperature. Avoid wetting the test strip holder.
Wipe dry or allow to air dry. Oily substances must be removed prior
to disinfection. Discard used towel in a sealed container where it will
not be touched by others.
6. Wash hands with soap and water.