MAGS1 User Guide
5
V1.5 22-1-2016
Sensor grounding with earthing rings:
2.2. Dry liner
Flowmeters with a hard rubber liner can show incorrect readings during the first 2-3 days after installation,
this is due to the fact that the time needed for transport and the time before installation is long enough for
the liner to dry out and thus it changes shape/size, this change effects reading accuracy.
Simply by keeping the meter wet, this problem will solve itself within 2-3 days and no other action is
required at all.
2.3. Electrical installation
Meter is produced with the cable already attached, in case of exchanging the cable pull the connection
cable through the cable gland on the top of the sensor, connect it according to color coding below.
2.4. Cables connections
The following diagram shows the connections of the cables between sensor and your RS485 bus and
power supply.
Recommended cable UNITRONIC® LiYCY (TP) 0035 830, 2x2x0.5.
Electrical installation should only be performed by a qualified person. Standard safety regulations
for hazardous electrical installations have to be respected.