Autofirm mode

2

1

Activate:

 Press and hold 

the button.

Comfort indicators flash 

and mattress inflates to 

Autofirm.

Extend:

Press button to extend 5 

minutes. Max. 30 min.

If not extended, pump 

goes back to previous 

mode.

1

3

Deactivate:

 Press and 

hold the button or select 

another mode.

Pressure reached –

Autofirm and Control 

comfort indicators turn to 

constant.

Transport mode (mattresses only)

2

1

Activate:

Press side buttons simultaneously and remove tubeset. 

The mattress supports the patient for up to 12 hours.

Deactivate:

Reconnect the tubeset. The pump goes back to the 

previous mode

CPR rapid deflation unit

Activate:

 Press both CPR release 

buttons simultaneously.

Reset: 

Press down the open part 

until it clicks into closed position.

“Click”

Make sure the front part is open.
The air evacuates in 15 seconds.

“Click”

 2s

2

Alternating mode, default setting

2

1

Activate:

 

Press button, the Alternating mode indicator is shown. 

Adjacent cells inflate and deflate on a 10 minute cycle.

Deactivate:

Select Reactive CLP or Autofirm mode.

If no further therapy is required, remove patient from 

the surface and turn off the pump.

Reactive CLP mode

2

1

Activate:

Press and hold the button. 
Reactive CLP mode indicator is shown. 

Deactivate:

Select Alternating mode or Autofirm mode. 

If no further therapy is required, remove patient from 

the surface and turn off the pump.