2s
2s
Autofirm mode
2
1
Activate:
Press and hold
the button.
Comfort indicators flash
and mattress inflates to
Autofirm.
Extend:
Press button to extend 5
minutes. Max. 30 min.
If not extended, pump
goes back to previous
mode.
1
3
Deactivate:
Press and
hold the button or select
another mode.
Pressure reached –
Autofirm and Control
comfort indicators turn to
constant.
Transport mode (mattresses only)
2
1
Activate:
Press side buttons simultaneously and remove tubeset.
The mattress supports the patient for up to 12 hours.
Deactivate:
Reconnect the tubeset. The pump goes back to the
previous mode
CPR rapid deflation unit
Activate:
Press both CPR release
buttons simultaneously.
Reset:
Press down the open part
until it clicks into closed position.
“Click”
Make sure the front part is open.
The air evacuates in 15 seconds.
“Click”
2s
2
Alternating mode, default setting
2
1
Activate:
Press button, the Alternating mode indicator is shown.
Adjacent cells inflate and deflate on a 10 minute cycle.
Deactivate:
Select Reactive CLP or Autofirm mode.
If no further therapy is required, remove patient from
the surface and turn off the pump.
Reactive CLP mode
2
1
Activate:
Press and hold the button.
Reactive CLP mode indicator is shown.
Deactivate:
Select Alternating mode or Autofirm mode.
If no further therapy is required, remove patient from
the surface and turn off the pump.