Preventive maintenance schedule
Preventive Maintenance for the AtmosAir™ MRS consists of regular cleaning (see Care and Cleaningsection on previous page) and an overall system check-
out to be performed at the intervals described below.
All components must be cleaned, disinfected and inspected after each patient’s use and before
use by a new patient.
Always use standard precautions, treating all used equipment as contaminated. Institutions should follow local protocols
for cleaning and disinfection.
Daily Cleaning
The cover should be wiped daily with a mild soap and water solution.
Inspection / System Check-Out
Check each of the following before placing the AtmosAir™MRS with a new patient:
1.
Check mattress surface for tears or cracking; do not use if tears or cracks are present.
2.
Ensure mattress is free of stains and is not overly faded.
Troubleshooting
It is recommended that all sections of this manual be reviewed before troubleshooting any
AtmosAir™ Mattress Replacement System.
Do not attempt troubleshooting outside this manual or where the remedy recommends to contact a ArjoHuntleigh service representative. Any
unauthorized service, modification, alteration, or misuse may lead to serious injury and/or product damage and will void all applicable warranties.
SYMPTOM
POSSIBLE CAUSE
SOLUTION
Mattress too firm upon arrival.
Difference in altitude not sufficient to
open valves.
Apply weight to mattress to open valves.
Mattress cover too wrinkled upon removal
from shipping container.
Internal components have not accomodated
to environment. (This does not affect inflation
or function).
Let mattress accommodate for 24 hours. If
problem continues, contact ArjoHuntleigh for
assistance.
Mattress does not inflate or is not firm.
Tubing not connected properly.
Check tubing inside mattress for
loose connectors.
Tubing kinked.
Check tubing inside mattress for kinks.
Tubing disconnected.
Check tubing inside mattress for
possible disconnect.
Holes in or damage to SAT.
Check SAT for holes or damage, or contact
ArjoHuntleigh for assistance.