Select alternating pressure mode by pressing 
the static mode button. The yellow indicator is no 
longer lit.
Position the patient’s limbs and head as directed 
by your hospital’s prone position protocol.
Utilizing Wound Valve Technology, speci

fi

 c air 

cells can be de

fl

 ated to of

fl

 oad pressure from 

sensitive body areas, i.e. the cells under the toes 
and genital area. Refer to product Instructions for 
Usel for full instructions.
Regular checks should be made to make sure 
the patient is free from the build up of pressure 
especially on anatomically sensitive areas.

Adjust individual Wound Valves 

as required to achieve pressure 

of

fl

 oading at speci

fi

 c anatomical 

points. 
• De

fl

 ate only one cell in the torso 

section.
• De

fl

 ate only one cell in the calf/

heel section.
Do not de

fl

 ate any more cells in 

each area or it may affect the 

support of the patient during the 
normal alternating cycle.

Select Alternating 

Mode by pressing 

the static button 

once. 

The yellow indicator 

is no longer lit.

Comfort Control set 
as appropriate. 

Supporting the patient’s head, slowly turn the 
patient over into prone position.
The patient’s shoulders should be positioned over 
the 4th cell from the head end of the mattress. 
Make sure that lines and monitoring equipment 
do not become trapped underneath the patient’s 
body. 
Adjust the head position using pillows, foam or gel 
pads so that a comfortable posture is achieved 
without hyperextension. 

All Wound Valves pointing up 
(all cells in

fl

 ated) 

except for Head Section Wound 
Valves which should be open if 
head section de

fl

 ation is required.

Remains on 
static mode.

Comfort control 
remains on 

fi

 rm.

649935EN_02 • 04/2020 • Page 2/2