Clip Slings Application
12
Attaching the Sling to the DPS
Fig. 13
1
2
Insert the clip attachment over the lug on the DPS.
Secure the clip in place by pulling the sling down so
that the lug is in the top portion of the clip hole.
WARNING:
Make sure all clips are correctly
engaged. Failure to do so could result in patient
fall.
Method 1
- Straight Attachment
Fig. 14
Apply the leg clips of
the sling onto the lugs
so that they become
positioned vertically.
For most patients, the straight attachment of the leg
clips is recommended.
Method 2
- Crossed Legs Attachment
Fig. 15
Cross the legs pieces
of the sling when
attaching to the lugs.
If the patient is prone to kicking off the leg clip, the
crossed attachment of the leg clips shall be applied,
which will prohibit the clip from being kicked off.
Detaching the Sling from the DPS
1
2
Fig. 16
Pull the sling up to release the lock.
Remove the clip attachment from the lug on the
DPS.
Operating the DPS
Fig. 17
Note
: To ensure maximum patient comfort, do
not allow them to hold onto the spreader bar.
The DPS spreader bar is adjusted by rising or
lowering the tilt handle until the patient is in the
desired position.
Lying
Position
Seated
Position