![Arjo Carevo Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/arjo/carevo/carevo_instructions-for-use-manual_2972254035.webp)
35
Always Follow these 24 Steps for Proper
Cleaning and Disinfection
If an Arjo shower panel will be used, see respective
IFU
.
Mechanical cleaning (Step 1- 8 of 24)
1 Put on protective gloves and eye wear.
(See Fig. 1)
2 Make sure the disinfectant is mixed according to the
instructions on the disinfectant bottle label or according to
the mix ratio stated in the shower panel
IFU
.
(See Fig. 2)
3 Remove the lower drainage and the pillow.
4 Rinse the
Carevo
shower trolley with water. Start from the
top of the mattress down towards the castors.
(See Fig. 3)
5 Scrub the
Carevo
shower trolley with disinfectant using a
soft bristle cleaning brush to remove contamination. Start
from the top of the mattress moving down towards the
castors.
6 Use a bottle brush soaked in disinfectant to clean the upper
drainage.
7 Remember to clean the underside of mattress, support plates
and outside of the upper drainage.
8 Using a soft bristle cleaning brush, scrub the pillow and the
lower drainage separately. Avoid contact with the mattress
until the pillow has been disinfected.
Continue with the steps on the next page.
Fig. 1
Fig. 2
Fig. 3
Содержание Carevo
Страница 1: ...04 BA 08_13EN 06 2019 Carevo 04 BA 08_13EN 06 2019 INSTRUCTIONS FOR USE ...
Страница 11: ...11 3 8 7 9 23 15 1 2 6 12 21 8 18 25 17 19 4 5 24 10 22 2 20 11 13 16 14 26 ...
Страница 31: ...31 Intentionally left blank ...
Страница 33: ...33 Fig 1 ...
Страница 37: ...37 Intentionally left blank ...
Страница 51: ...51 Intentionally left blank ...