2880R-LT, 2881R-LT, 2882R-LT
Revision 1.0
P a g e 16 | 28
6.0 STARTUP AND CALIBRATION
6.1
Startup
Check that the power wiring, PMC wiring and probe are wired in accordance with the electrical
installation drawing. Connect the probe input terminal block TB6 as per figure 3.
Power On the unit. Allow a 1 hour warm-up period before calibrating.
The Status LED on the controller circuit board should be green. A red Status LED indicates a fault
condition. If red, check the Troubleshooting section.
The LCD will scroll to the normal operating display. See section 10.0 Menu Flow Chart.
NOTICE
An instrument setup and field calibration is required on initial power
up. See section 6.3 for controller setup and section 6.4 to calibrate the
transmitter.
6.2
Menu Flow Chart Back Chart Background Information
The control setup, diagnostics, and calibration are accessed using the display and keypad on the
controller. The Flow Chart in Section 10.0 provides an overview to the various menus and features.
Keep a copy of the flow chart at hand when accessing the internal controller features.
Below is a description of the menu functions.
6.2.1
Menu Short Form
Since the 288XR controller has a small LCD, some menu description may be in short form. The
following are the menu descriptions:
Short Form
Description
Short Form
Description
Diags:
Diagnostics
SENSTVTY:
Sensitivity
Cal Pts:
Calibration points
A1:
Alarm relay 1
Auto Cal:
Auto calibrate
A2:
Alarm relay 2
Man Cal:
Manual calibrate
Alrm: Alarm
Cal Ok:
Calibrate ok
Alrm Lvl:
Alarm level
Cal Err:
Calibrate err
HiLv1:
Hi Differential alarm set value
mA out:
mA output
LoLv1:
Lo Differential alarm set value
mA Span:
mA output span
Delay: Alarm
Delay
Sec: Seconds
^SP:
Relay Setpoint Hi action
ALRM CAL:
Alarm Calibration
vSP:
Relay Setpoint Low action
Cal: Calibration
point
Dir:
Relay Action (Dir = Direct)
Net ID
Network Identification
Fsafe: Failsafe