AZI Customer Service 800-528-7411 or 602-470-1414
Page 18 of 57
SET INTERVAL
If auto-sampling is being used (as set in
SAMPLE MODE
above), the J605 can be configured to
automatically perform sensor re
generations at a predetermined interval. Use the ▲and ▼arrows
to scroll through the available intervals of 6, 12, 24, and 48 hours, and press
ENTER/START
to
save the selected interval. The J605 will now automatically perform a sample followed by a
regeneration every 6, 12, 24, or 48 hours as selected, starting at the time set in
SET HOUR
below. Use
ESC
to exit without saving.
SET HOUR
Use the ▲and ▼arrow keys to select the hour of the day (using a 24-hr clock) to start
regeneration. Press
ENTER/START
to save the setting. At the time set, if auto-sampling is
being used, the J605 will automatically take one sample and then perform a regeneration. This
process will repeat at the interval indicated in
SET INTERVAL
above. Use
ESC
to exit without
saving.
NOTE: If the sensor is saturated and a regeneration cycle is performed and completed before the
set hour, the instrument will perform another regeneration at the set hour. However, if the
regeneration cycle does not complete prior to the set hour, the regeneration at the set hour will be
skipped.
DATA menu
Sample data can be stored within the instrument, up to a maximum of twenty thousand (20,000)
data points. On instruments that have the USB Communications option, in addition to viewing
the sample data on the instrument, data can be retrieved from the instrument in one of three ways:
1.
SEND TO PC
2.
SAVE TO USB MEMORY DRIVE
3.
PRINT DIRECTLY TO AN AZI APPROVED PRINTER (AZI P/N: Y990-0098)
The presence of the USB icon on the main screen indicates that the Communications option has
been purchased and that the USB ports are functioning properly. The USB icon will flash during
data transmission. See
for the location of the USB icon.
The data is stored in non-volatile memory and will not be affected if the battery is changed or
disconnected.
VIEW STORED DATA
Select
VIEW STORED DATA
to view previous sample and regeneration history on the screen
of the J605. The Location Name (if specified) corresponding to the data is displayed at the top of
the screen and three sample values at a time are displayed at the bottom of the screen, along with
the time and date the sample was taken. The most recent sample is displayed at the top of the list,
and regenerations are indicated by the word “Regen” instead of a sample value.
Use the ▲ and
▼ arrow keys to scroll through the data. If Location Names are used, the Location Name
displayed will automatically update during scrolling to reflect the sampling location of the data
being displayed. The J605 will beep-beep upon reaching the end of the list. After viewing the
data, press
ESC
to exit and return to the
DATA
menu.