26 ARGUS settings
ARGUS 152
357
Daylight
savings time:
If central European time is selected, ARGUS calculates with a
total time offset of UTC +2 during the summer months.
Date / time:
To set the time manually, enter the correct time in your time
zone using the ARGUS number keys. In automatic time set-
ting, ARGUS automatically reads the time from a preconfig-
ured time server.
Default:
0.de.pool.ntp.org
This can be changed to a different server. The prerequisite is
that ARGUS can connect to the internet. Carry out e.g. a ping
test (e.g. ping www.argus.info).
The time you enter runs on the built-in ARGUS realtime clock until the power
supply is interrupted. When ARGUS is switched off and the batteries
removed, the clock continues to run for a few days on its internal buffer. The
time is undefined as soon as the buffer is exhausted, and must be set again.
Management
interface
Start
management
interface
Determines whether the management interface is used. If
WLAN was selected as the management interface, ARGUS
operates as a WLAN router, see page 358.
Depending on the selected interface, ARGUS displays either
a WLAN or an Ethernet symbol.
WLAN:
WLAN is not active (gray)
WLAN is active (green)
Ethernet:
Ethernet is selected
Default:
off
VNC server
Start VNC Determines whether the VNC server is used.
Default:
off
VNC
scaling
Determines the scaling used to display the
ARGUS screen on the PC.
Range: Factor 1 - Factor 4
Default:
Factor 2
WLAN
When WLAN is grayed out, the management interface must
first be halted (see above) by setting "Start management inter-
face" to off. Then restart it again.
SSID
The service set identifier (SSID) is a user-defin-
able name with which ARGUS can be identified as
a network. Your SSID can also be displayed using
the keyboard shortcut
and
Default:
Argus152_SerialNumber