12/17
5084132-03
Verify if
calibration
of the appliance sensor is re-
quired.
We recommend to calibrate the appliance sensor if a
temperature deviation
of more than
0,5°C
between
the test probe placed in the center of the appliance
and the temperature shown on the display is found
(customers own quality standards may define other
requirements according to calibration, these shall
then be valid). The average value of the two center
probes in 1400 models shall be used as the base of
the calibration of 1400 models.
If calibration is performed, note the calibration value
and repeat above steps.
Calibration parameter:
/c1
Calibration requires the following field to be signed
to confirm calibration has been performed:
Date:
Signature:
Temperature of test
Probe (center of appli-
ance):
Display temperature:
Temperature deviation:
Note setting of calibration
parameter:
(if calibrated)
Verify the
temperature deviation
of the appliance by
comparing the maximum and minimum temperature
measured during a period of min.
2 hours
.
Approval criteria
Temperature deviation of max. 5K between the
warmest and coldest temperature value measured.
Probe 1
Max. temp.:
Min. temp.:
Probe 2
Max. temp.:
Min. temp.:
Probe 3
Max. temp.:
Min. temp.:
Probe 4
Max. temp.:
Min. temp.:
Probe 5
Max. temp.:
Min. temp.:
Probe 6
Max. temp.:
Min. temp.:
Max. temp. deviation (K):