LRE 700/1400 & LFE 700/1400 

 

IQ/OQ

 - INSTALLATION AND 

OPERATIONAL QUALIFICATION 

Содержание LFE 1400

Страница 1: ...LRE 700 1400 LFE 700 1400 IQ OQ INSTALLATION AND OPERATIONAL QUALIFICATION...

Страница 2: ...intended to verify that the appliance is delivered in corre spondence with the ordered appliance and is installed in a correct environment The Operational Qualification is intended to verify the func...

Страница 3: ...e check points 5 2 3 Environmental verification 7 3 Operational Qualification 8 3 1 Parameter verification and adjustment 8 3 2 Placement of test probes 9 3 3 Temperature mapping 11 3 4 Alarm and temp...

Страница 4: ...____________________ Date of qualification DD MM YYYY _________________________ Qualification carried out by name _________________________ Customer _________________________ Location of the appliance...

Страница 5: ...shall fill out the signature list below in order to establish traceability to the test signatures All devices used for the execution of the current IQ OQ shall be noted in below device list SIGNATURE...

Страница 6: ...verify the following points DOCUMENTATION CHECK COMMENTS Ensure operating instruction manual has been supplied with the appliance Ensure operating instruction manual has been read and un derstood by t...

Страница 7: ...ped with levelling feet Controller Ensure no scratches on the display Ensure that all cables are mounted according to safety requirements Remote alarm connection Ensure that the remote alarm contact i...

Страница 8: ...nsure no direct sunlight on the appliance Ensure that the appliance is not located near heat sources Ensure that the room is well ventilated Surface Ensure installation on a solid and flat surface to...

Страница 9: ...E THE VALUE COMMENTS Password PS 22 22 Set point St 4 20 Control delta hysteresis rd 2 2 Interval between defrost dl N A 6 hrs High tempera ture alarm by pass time after defrost and door open d8 0 0 A...

Страница 10: ...illustration see the red arrows Each test probe must be loaded with a 25 gr brass block or equivalent to ensure precise temperature measurement LRE LFE 700 PROBE LOCATION Probe no 1 45 cm distance fro...

Страница 11: ...Probe no 3 45 cm distance from the bottom of the interior cabinet left side of cabinet Probe no 4 45 cm distance from the top of the interior cabinet right side of cabinet Probe no 5 72 cm distance f...

Страница 12: ...on of the appliance by making sure that the temperature of the appliance equals the ambient temperature Start the temperature logging of the test probes Turn on the appliance and let it run until stab...

Страница 13: ...he calibration value and repeat above steps Calibration parameter c1 Calibration requires the following field to be signed to confirm calibration has been performed Date Signature Temperature of test...

Страница 14: ...of the door open alarm bypass parameter d8d is 2 min Open the door and note start time and start tem perature Leave the door open until the door open alarm activates acoustic alarm and visual error co...

Страница 15: ...rature value within the normal operating tempera ture set point hystere sis Temperature of test probe Recovery time Time of door closing time of reaching a temper ature value within the nor mal operat...

Страница 16: ...he AL parameter is a relative threshold to the set point see examples in point 3 1 Approval criteria Refrigerator Activation of the acoustic and visu al LO alarm according to the AL low temp alarm Ad...

Страница 17: ...ctive action has been initiated and or finalized In case of deviations which cannot be corrected on site please contact the distributor of the appliance for support in correction of the deviation s DE...

Страница 18: ...4 Result of Installation and Operational Qualification PRODUCT INFORMATION Manufacturer Model Serial number RESULT CHECK Passed Failed CHECKED TESTED BY DATE SIGNATURE AND STAMP VERIFIED BY DATE SIGN...

Страница 19: ...5084132 03...