Arctiko BBR100 Скачать руководство пользователя страница 1

OPERATING INSTRUCTIONS

BLOOD BANK REFRIGERATOR
100/300/500/700/1400

Содержание BBR100

Страница 1: ...OPERATING INSTRUCTIONS BLOOD BANK REFRIGERATOR 100 300 500 700 1400...

Страница 2: ...er 12 Location of Refrigerator 13 Installation 14 Control Panel 15 Description Control Panel for user 15 Description Control Panel for engineer 16 Default settings 17 Operation 18 Chart Recorder 19 Pr...

Страница 3: ...od Bank unit is designed for storage of whole blood and related products at constant temperature in clinical pharmaceutical research and other laboratory applications Please carefully read this manual...

Страница 4: ...e explosion or fire Never install the unit where acid or corrosive gases are present as current leakage or electric shock may result due to corrosion Always ground earth the unit to prevent electric s...

Страница 5: ...step on the power supply cord or never damage or break the power supply plug A broken supply cord or plug may cause fire or electric shock Do not use the supply cord if its plug is loose Such supply c...

Страница 6: ...to tip over the unit during movement to prevent damage or injury Prepare a safety check sheet when you request any repair or maintenance for the safety of service personnel Always keep the keys in se...

Страница 7: ...at least 16cm free at the back The appliance meets current EU directives Low voltage 2006 95 EEC Electromagnetic compatibility 2004 108 EEC Only use the appliance for its specific purpose The applian...

Страница 8: ...or Parts and their Functions 1 Control panel 2 Light Power Switch 3 Battery Back up 4 Door lock with key 5 Door Handle 6 Castors 7 Chart Recorder 8 Blood Storage Drawers 9 Interior Light 8 Adjustable...

Страница 9: ...easy of opening and closing door 6 Castors The 4 castors underneath the unit provide the convenience of moving the unit around 7 Automatic Temperature Recorder The refrigerator is equipped with an aut...

Страница 10: ...Before you start the refrigirator and freezer you have to connect the battery backup 10...

Страница 11: ...Mount a screw inside the Handle both in top and bottom 1 Door Handle with Fittings and Cover 2 Place the Fittings on the Door Handle 3 Mount the Handle on the door with screws 4 Push the cover for th...

Страница 12: ...How to mount the Adjustable Devices in the Blood Storage Drawer Place the device in both split in the Drawer Push the device in its place 12...

Страница 13: ...ooling capability 7 Ground protection Grounding the refrigerator will prevent electrical shock to operators if its electrical insulation become weak 8 Do not place any objects on top of the refrigerat...

Страница 14: ...lves to see if the liquid level is up to the 200 mark inside the bottle If the bottle has no fluid or the fluid level is lower than the 200 mark use the following steps to fill the bottles at the uppe...

Страница 15: ...rature too high for longer time than alarm delay 7 Low temp alarm Temperature too low for longer time than alarm delay 8 Door op alarm If the door is left open the door alarm will be activated 9 Probe...

Страница 16: ...t 14 Max temp limit Max set point allowed to be set on level I 15 Min temp limit Min set point allowed to be set on level II 16 Password protected Yes no Password for start stop the freeer Yes or No 1...

Страница 17: ...gerator 2 Door open alarm Refrigerator 01 Probe eprom failure Refrigerator 01 Alarm Delay Refrigerator 01 Alarm log Refrigerator 01 Description Arctiko default settings Customer setting LEVEL II Passw...

Страница 18: ...is too high 6 Inspect air channels to see if they are blocked 7 Check the condenser air filter to see if it is clogged with dirt 1 Make sure the unit is placed on a level floor 2 Make sure any part o...

Страница 19: ...pen inscription The case front panel is a key locked transparent polycarbonate door The diagram drive is ensured by A standard quartz two speed clockwork mechanism 1x24h and 7x24h powered by a standa...

Страница 20: ...t pen in place Lift the plate s pointer 2 Insert the end of the pointer into the fibre point pen s slide rail up to the stop 6 Remove the cap by pulling and turning it at the same time Do not place th...

Страница 21: ...so as to obtain an accurate time setting Tighten locking screw without touching the diagram up to positioning the diagram on the exact time The taking up of the angular play shall be carried out at th...

Страница 22: ...re supplied with the unit usually last about half a year When you are running short of the chart paper please contact Arctiko to purchase more The lead time to deliver the charts is 1 3 working days a...

Страница 23: ...adjust the temperature set point 5 It usually takes several hours for the unit to reach the preset temperature Once the unit s temperature is stable check the sensor bottle temperature to see if it ma...

Страница 24: ...urfaces 4 Do not pour water directly into the refrigerator By doing so the water can damage the insulation materials and cause problems 5 Parts in the refrigeration system for this refrigerator are co...

Страница 25: ...e Unplug the refrigerator from the power supply Remove the cover for the fluorescent light tube Turn the tube and pull it out the split Push and turn the new tube in its place Install the cover again...

Страница 26: ...condenser Finned coil evaporator R 134a Automatically Forced defrost system Electronic Controller Varnished steel Stainless Steel Electric heater heated glass door Model Number 26 BBR100 BBR300 BBR500...

Страница 27: ...Non Volatile memory storage 8 W DM8 6 8V6AH Standard including one box of recorder chart paper and 1 piece of 9 V battery One set of key one certificate of warranty one copy of operation manual one pl...

Страница 28: ...current A 6 Voltage 7 Symbol for Rosh 8 CE sign for Declaration and Conformity 9 Made in EU 10 EN standard 11 Refrigerant and quantity 12 Climatic class 13 Inside temp C 14 Weight kg 15 Serial no 30 w...

Страница 29: ...at you can easily fit yourself please give unit type number and the function of the part s concerned If any spare parts are required we will try to have a day to day delivery service for spare parts b...

Страница 30: ...BBR Units Block Diagram of Main PCB board 30...

Страница 31: ...Wiring diagram for BBR 500 700 ST 31...

Страница 32: ...Wiring diagram for BBR 1400 ST 32...

Страница 33: ...Notes 33...

Страница 34: ...Notes 34...

Страница 35: ...Notes 35...

Страница 36: ...50 80 034 04 GB WWW ARCTIKO COM LAMMEFJORDSVEJ 5 DK 6715 ESBJERG N DENMARK TEL 45 70 20 03 28 FAX 45 70 20 03 29 INFO ARCTIKO COM WWW ARC TIKO COM...