X-Ray & ABS
22
Pulsed Boost
WARNING:
Steps within this procedure produce x-rays. Use appropriate precautions.
1. With BOOST enabled, select PULSE on the control panel.
2. Select AUTO Fluoro mode and take an exposure. Once the image has stabilized, take another by pressing on the BOOST
footswitch.
3. Verify that the audible alarm beeps at twice the normal rate.
4. Verify that the mA elevates until the image becomes stable.
Verify Fluoro Timer and Timer Test
1. In FLUORO mode, the accumulated exposure time is displayed in the center of the display panel. After five minutes of
accumulated fluoro x-rays, the timer alarm will sound.
2. Press ALARM RESET to reset the alarm. The alarm will stop, but the accumulated exposure display will not reset.
3. Press and hold the ALARM RESET for five seconds. The accumulated timer display will reset to zero (0).