“Measured Quantity” displays the value present at the Model 931A
measurement inputs in actual measurement units (e.g. volts, amps).
“Percent Full Scale” displays the value present at the Model 931A
measurement inputs as a percentage of full-scale. Full-scale is defined
by the lower and the upper limit of the currently selected input range
(see Section 3.8.1.3 for a description of Input Ranges).
3.8.1.6
Display Output As
Select Display Output As from the Transducer Configuration menu to
display the following sub-menu:
“
Measured Quantity
” sets the Model 931A to display the actual
value measured at the output of the transducer under test (i.e., volts
or milliamperes).
“
Percent Full Scale
” displays the measured value (at the output
of the transducer under test) as a percentage of full-scale. Full-scale
is defined by the lower and the upper limit of the selected output
range. Does not apply for Custom Pulse.
“
Percent Error
” displays the measured error of the transducer
based on the measured input and output quantities.
To properly utilize the Percent Error feature, enter the following: (1)
the upper and lower transducer input quantities and (2) the upper
and lower transducer output range limits. Setting the output range
limits are critical for transducers having a lower output range limit of
other than zero. For example a typical transducer may have a 4
−
20
mA output. With such a transducer, zero at the input corresponds
to 4 mA at the output.
69
Содержание 931A
Страница 6: ...vi ...
Страница 12: ...xii ...
Страница 14: ...xiv ...
Страница 16: ...xvi ...
Страница 52: ...SOURCE LOAD B A C N Figure 3 12 3 Phase 4 Wire 3 Element Connections 36 ...
Страница 90: ...Iout I1 V1 V2 V3 I2 I3 Figure 3 17 3 Phase 4 Wire 3 Element Transducer Test 74 ...
Страница 114: ...98 ...
Страница 120: ...104 ...
Страница 142: ...126 ...