INSTRUMENT DEFAULT SETUP
EXIT
(Nothing Disturbed)
PARTIAL
(Calibration Not Disturbed)
FULL
(All Values Set To Default)
Table 2.4:
931A Reset Selection Screen
2.8.2.1
Resetting Procedure
NOTE: DO NOT SELECT “FULL” UNLESS YOU INTEND TO
COMPLETELY ERASE THE INSTRUMENT CALIBRATION!
NOTE: Normally, select “PARTIAL” to restore the instrument
to original default conditions.
1. Press the ON/OFF button to start the unit.
2. Press and release the ENTER key when the first display screen
becomes visible.
3. Using the Cursor keys (up or down arrows) select “PARTIAL” from
the three options shown in Table 2.4.
4. Press ENTER and the instrument should ask, “ARE YOU SURE??
YES=UP NO=DOWN”. Press the UP arrow to complete the reset
procedure. Otherwise, press the DOWN arrow to escape.
5. The display should indicate the message, “PLEASE WAIT 10s TO
4m”. After a short time the display should return to the final VOLTS
AMPS display indication.
6. You may need to configure your settings under MENU and/or SHIFT
>
SCALE.