background image

10

A/S Inquiry : +82-2-3452-8966 / [email protected]

USER MANUAL

MINI CENTRIFUGE

6. Troubleshooting

The device can not be started

1. Power failure

    → Check the mains connection.

2. AC adaptor defect

   → Check the output DC volt of the adapter using a multimeter

  3. Lid sensor failure

  → Contact Hanil Technical Support Team.

4. If the beep does not sound right after connecting the power supply, 

  → Contact Hanil Technical Support Team.

      +82-2-3452-8966 / [email protected]

Содержание Smart 6

Страница 1: ...ARA Smart 6...

Страница 2: ...ARA Smart 6 Research Use Only Copyright 2019 Hanil Scientific Inc All rights reserved Contact Us If you have any questions contact Hanil Scientific Inc techsupport ihanil com info ihanil com...

Страница 3: ...tended Use 5 3 Product overview 6 4 Technical specifications 7 5 Installaion Operation 8 5 1 Power connection 8 5 2 Operation 9 6 Troubleshooting 10 7 Rotor 11 CONTENTS The appearance or specification...

Страница 4: ...2 According to IEC61010 2 020 maintain a 30cm clearance envelope around the device while the rotor is spinning 3 Do not operate the device in areas where work is completed with explosice substances 4...

Страница 5: ...nd contact Hanil Technical Support Team 8 Do not block vents 9 Do not drop or subject the device to severe impacts 10 Caution when handling the AC DC adaptor Do not bend or pull the AC DC adaptor Use...

Страница 6: ...pen button Power Connector Running Mode Selection Switch Operation mode switch Lid Closing the lid Start the centrifugation Opening the lid Stop the centrifugation Short Mode S Stop the centrifugation...

Страница 7: ...Rotor 1 5 2 0 mL x 6 hole Rotor 2 PCR Strips 0 2 mL 0 5 mL Adaptor Optional Max RPM 6 500 RPM Max RCF Angle Rotor 1 5 2 0 mL x 6 hole 2 234 xg 2 PCR Strips Rotor 1 412 xg Radius Angle Rotor 1 5 2 0 mL...

Страница 8: ...adaptor Keep this manual near the device at all times 1 Connect the power cable to the power socket on the back side of the device and plug the cord into the outlet 2 A beep sounds when the power is...

Страница 9: ...s the lid open button 3 Opening the lid 2 Starting Stopping the centrifugation Short Mode S When the lid is closed centrifutaion starts Operataion stops after 12 sec of operation Continuous Mode C Ope...

Страница 10: ...ed 1 Power failure Check the mains connection 2 AC adaptor defect Check the output DC volt of the adapter using a multimeter 3 Lid sensor failure Contact Hanil Technical Support Team 4 If the beep doe...

Страница 11: ...A2 0 6 Hole angle 40 Max Capacity 6 x 1 5 2 0 mL Size x H 88 x 24 5 mm 1 5 2 0 11 47 3 6 500 2 234 0 5 TR0 5 2 8 x 31 36 2 6 500 1 710 0 2 TR0 2 2 6 5 x 23 29 2 6 500 1 379 A0 2 16 Hole angle 40 Max...

Страница 12: ...im ara com OMM7PDEN1902 C...

Отзывы: