Panel Mounting
Panel adapter
Panel adapter
Front protective lid
Unit:mm
E
.
OPTIONAL PARTS DIMENSIONS
I.
ZERO POINT
RESET
Under measure mode, press button
for 1 ~ 3 seconds.
H.
PRESSURE UNIT SETTING MODE
Press button for more than 3 seconds.
Measure mode
Use button to set desired pressure unit.
Unit setting
Measure mode
MPa
MPa
MPa
Measure mode
MPa
Measure mode
MPa
Zero point setting
MPa
MPa
F.
PRESSURE UNIT CONVERSION TABLE
J.
DUST AND SPLASH
ENVIRONMENT
To maintain IP65, please insert an air
tube (O.D.: 4mm) into atmospheric
release port and pipe the other end away
from dust / spattering liquid environment.
Use pr
oper air tube dimension of
O.
D.: 4mm, I.D.: 2.5mm, and the length
is de
pending on
the user’s environment.
Atmospheric release port
●
●
Air tube
Do not press button for 3 seconds to return
to measure mode.
Release the button when “ ” is shown.
K.
ERROR CODE INSTRUCTION
Residual
pressure error
During zero reset, ambient
pressure is over
±3% F.S.
Change input pressure to ambient
pressure and perform zero reset again.
Applied
pressure error
Supply pressure exceeds the
upper limit of pressure setting.
Adjust the pressure within operating
pressure range.
Error Type Error code
Error Condition
Troubleshooting
Battery
error
When battery level is under
2.40V, the battery symbol
and will flash 4 times
then enter power save mode.
Replace battery.
From
To
1 MPa
0.006895
0.07031
1
0.06895
10
145.038
10.197
1
0.100000
1.01972
14.5038
1
0.980665
14.2233
1
0.0980665
1 kgf/cm
2
1 psi
1 bar
MPa
kgf/cm
2
psi
bar
G.
POWER-SAVE MODE
Measure mode
Power-Save mode
MPa
◎
The pressure gauge will enter power saving mode if the button is not
pressed within 60 sec.
No button operation for 60 seconds
Panel adapter
Front protective lid
Panel adapter
PA-D
MPa
kgf
bar
psi
PAP