Model 5000A225C
36
Rev A
2.8.6.23
Clear Fault Log
Use this command to clear the entire contents of the error log. This action is permanent and cannot be undone.
Syntax:
FSTA:LOG:CLEAR
Response Format:
None
Example:
To clear the entire fault log, send the following command:
FSTA:LOG:CLEAR<LF>
2.8.6.24
Remote Operating Timeout (ROPTO)
Due to the high-power output capability of this amplifier, there is a built-in safety mechanism while the Keylock
switch is in the REMOTE position. Once the RF:ON command is issued, it is necessary to continually
communicate over any of the remote I/O ports on a periodic basis.
This command sets and queries the remote operating timeout value, in seconds.
Syntax:
ROPTO<timeout>
<timeout>:
The seconds (between 0 and 500) that the amplifier will allow the RF to stay
enabled before requiring a remote command. If exceeded, the RF will be
disabled.
Query:
ROPTO?<LF>
Response Format:
ROPTO=<timeout><LF>
Example 1:
To set the timeout value to 1 minute, send the following command:
ROPTO60<LF>
Example 2:
To query the timeout value, send the following command:
ROPTO?<LF>
Response:
ROPTO=6<LF>
(6 seconds of remote inactivity will disable the RF)
Note:
It is recommended to use the STATE? or FSTA? queries to avoid the timeout,
but any command or query will suffice.
2.8.6.25
ALC Board Firmware Query
This command queries the firmware of the ALC board assembly.
Query:
*ALC?
Response Format:
ALC_SW_REV<revision><LF>
<revision>:
The revision of the ALC Board.
Example:
To query the ALC Board firmware revision, send the following command:
*ALC?<LF>
Response:
ALC_SW_REV1.60<LF>
(Firmware is v1.60)