![AR 10046836 Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/ar/10046836/10046836_operating-and-service-manual_2962294035.webp)
Model 500W1000C
Rev A
15
2.6.3
Fiber-Optic Communication
The Fiber-Optic port is a serial communications bus. All commands and queries through this port must be
terminated with a Line Feed character. When a valid query is received, it is processed and the result is
immediately transmitted back over the Fiber-Optic interface. This port is designed to time-out if there is no
activity on the bus for more than 5 seconds. At this time the internal buffer is cleared and a
TIMEOUT_ERROR
message followed by a Line Feed is sent out from this port.
The baud rate for the Fiber-Optic Serial port is user selectable from the RS-232/Fiber-Optic Serial Baud Rate
selection screen. To get there from the Main Menu, touch the User menu button followed by the I/O menu
button and finally the Serial menu button. A thin black outline indicates the present baud rate selection. When
the back button is pushed the baud rate selection is stored to non-volatile memory. If power is lost prior to
hitting the back button, any baud rate selection changes will be lost.
NOTE: This baud rate setting is shared by both the Fiber-Optic serial port and the RS-232 port.
The Fiber-Optic port provides the user with the ability to optically isolate the controlling PC from the amplifier.
This can be useful in an environment where RF/Microwave energy could be coupled onto a connection to one
of the “wired” communications ports and fed back to the controlling PC.
Both optical connections (TX and RX) are optimized to work with light at a wavelength of 820nm. For more
detailed specifications on this port, consult the Avago HFBR series datasheet found at
A glass, multi-mode, fiber-optic cable of 200um is recommended, however fiber-optic cable as small as 50um
can be used. The connector type for this port is ST.
This port can be used in conjunction with either an AR model IF7000 RS-232 to Fiber-Optic Interface (1200 to
9600 baud only) or an AR model IF7001 USB to Fiber-Optic Interface (19200 baud only). Note that these
devices use SMA connectors so a fiber-optic cable is needed with ST connectors on one end and SMA
connectors on the other. This cable can be obtained from a fiber-optic cable distributor such as Fiber Instrument
Sales (FIS). Their web-site can be found at
. An example cable that will work
for this connection is FIS Part Number D615M7FIS. The 7 in the part number refers to the length of the cable.
In this case the length is 7 meters.
Table 2-3. Fiber-Optic Serial Port Settings
Word Length
8 bits
Stop Bits:
1
Baud Rate:
User selectable (default is 19200)
Parity:
None
HW Handshake: