Model 175A400 

Rev C 

13 

If the connected network does not have DHCP enabled then the device can be assigned an IP address by the 

user.  To do this, download the utility called DeviceInstaller

TM

  from 

www.Lantronix.com

.  For assistance 

using this utility please consult the utilities embedded help file. 

The DeviceInstaller

TM

 utility will scan the network and find all connected Lantronix Ethernet devices.  This 

list of found devices will include any connected AR Ethernet devices.  By selecting one of the connected 

devices from the list, its IP address and subnet mask can be changed along with a number of other settings.  

One should use caution in adjusting any settings he/she is unfamiliar with as doing so may cause the port to 

become unresponsive.  By default the port for the TCP data channel is 10001. 

*DHCP is a protocol used to assign a dynamic IP address to a device.  Network server software will assign an 

available IP address to a device when it is connected and powered on.  Different IP addresses may be assigned 

at different times. 

2.4.6 

Remote Commands 

 

If a command or query is unrecognized it is echoed back out the port it came in on. 

 

All commands and queries are terminated with a Line Feed character. 

 

A Line Feed character is indicated by <LF> in subsequent command and query definitions. 

 

All queries can be sent when the Toggle switch is in the LOCAL or REMOTE position. 

 

All commands can only be sent when the Toggle switch is in the REMOTE position. 

 

All spaces in command and query definitions are indicated by <space>. 

 

If a query is recognized, its associated response is echoed out the port it came in on. 

A COMMUNICATIONS_ERROR<LF> can occur if the time between commands or queries is too short, or 

the internal RS-485 link between the IO Board assembly and the Multipurpose Board (MPB) assembly is 

broken. 

The development of application programs requires an understanding of the operation of the 250A400 as well 

as the intended application. 

An application program on the computer/controller should issue only one character string (command or 

query) at a time.  After each functional command is issued, the 175A400’s status should be checked to ensure 

that the command has been properly executed.  The application program should allow sufficient time for the 

function to be completed before checking the status. 

The application program should facilitate the checking of the status just prior to issuing a command, since the 

status could have been changed by a fault condition or by operator actions. 

Variables represented by wild card characters i.e. x, y, z etc. do not indicate or delimit the number of 

characters actually specified. 

Содержание 10044831

Страница 1: ...Operating and Service Manual 175A400 Model 10044831 Part Number Serial Number 160 School House Road Souderton PA 18964 215 723 8181 Fax 866 859 0582 www arworld us...

Страница 2: ......

Страница 3: ...Requirements for Electrical Equipment for Measurement Control and Laboratory Use Part 1 General Requirements UL 61010 1 Issued 2012 05 11 Ed 3 Safety Requirements for Electrical Equipment for Measure...

Страница 4: ......

Страница 5: ...n folgenden Minimalwerte der Schirmd mpfung und Filterung in den unterschiedlichen Leistungsklassen empfohlen Hochfrequenzleistung min Schirmd mpfung 100 Watt 50 dB 101 1000 Watt 60 dB 1001 10 000 Wat...

Страница 6: ......

Страница 7: ...clearance on all vented sides of the equipment to permit the airflow required for proper ventilation Do not restrict airflow into the equipment by blocking any vents or air intakes Restricting airflow...

Страница 8: ...a assembly may be heavy requiring two persons to lift Use caution when installing or removing unit Follow all equipment setup and lifting instructions specified in this document Ensure connectors are...

Страница 9: ...von 10 2 cm 4 in auf allen bel fteten Seiten des Ger ts ein um eine ausreichende Luftzirkulation zu gew hrleisten Beintr chtigen Sie den Luftstrom des Ger ts nicht indem Sie L ftungs ffnungen oder de...

Страница 10: ...allen in diesem Dokument angegebenen Anweisungen zur Anbringung Stellen Sie sicher dass alle Anschl sse f r den beabsichtigten Betrieb geeignet sind Informationen zu den Anschl ssen erhalten Sie im Be...

Страница 11: ...tilation Ne limitez pas le flux d air allant dans l quipement en bloquant tout vent ou entr e d air La restriction du flux d air peut endommager l quipement causer des coupures intermittentes ou des d...

Страница 12: ...dent lorsque vous installez ou retirez l unit Respectez toutes les instructions concernant l installation et le levage de l quipement indiqu es dans ce document Assurez vous que les connecteurs sont a...

Страница 13: ...is voor goede ventilatie mogelijk te maken Belemmer de luchtstroom in het apparaat niet door ventilatieopeningen of luchtinlaten te blokkeren Het belemmeren van de luchtstroom kan leiden tot schade a...

Страница 14: ...nstellen en optillen van de apparatuur die in dit document worden vermeld Zorg ervoor dat de connectoren geschikt zijn voor de beoogde werking De connectoren worden gespecificeerd in de gebruikershand...

Страница 15: ...ace 10 2 4 1 GPIB IEEE 488 Communication 10 2 4 1 1 Setting the GPIB IEEE 488 Address 10 2 4 2 RS 232 Communication 10 2 4 3 Fiber Optic Serial Communication 11 2 4 4 USB Communication 12 2 4 5 Ethern...

Страница 16: ...Troubleshooting 28 4 3 1 General Reading Faults 28 4 3 2 The Unit Cannot Be Operated Remotely 29 4 3 3 Thermal Fault Schematic Diagram Nos 10044790 10039874 29 4 3 4 Interlock Fault Schematic Diagram...

Страница 17: ...djustment of the amplifier s RF Gain during CW mode operation and provides graphical displays of the amplifier s Forward and Reflected power levels A General Purpose Interface Bus GPIB IEEE 488 2 inte...

Страница 18: ...air into or exiting from the unit will not be restricted diverted or re circulated through the unit itself in particular the flow of warm air exiting the rear of the amplifier should not be impeded Do...

Страница 19: ...gh gain and linearity are re quired Available in a stylish contemporary cabinet for benchtop use or with cabinet re moved for rack mounting The Model 175A400 when used with a sweep generator will prov...

Страница 20: ...inus 20 dBc maximum at 150 watts SPURIOUS Minus 73 dBc typical PRIMARY POWER Universal selected automatically 100 240 VAC 50 60Hz 770 watts CONNECTORS RF Input N female RF Output N female REMOTES PACK...

Страница 21: ...Page 3 175A400 175 Watts CW 10kHz 400MHz Graphs...

Страница 22: ...Page 4 175A400 175 Watts CW 10kHz 400MHz Graphs...

Страница 23: ...er call 800 933 8181 Direct to Service call 215 723 0275 or email service arworld us For Faxing Orders 866 859 0582 Orders Only Please info arworld us Approved for public release by AR RF Microwave In...

Страница 24: ......

Страница 25: ...d to withstand without damage RF input power levels up to twenty 20 times its rated input of 1mW However signal levels higher than 20mW or transients with high peak voltages can damage the amplifier A...

Страница 26: ...ight emitting diode LED in the switch indicates whether the 175A400 power is on or off The main power supply fans are active when power is on The LCD touch display is active as long as the main circui...

Страница 27: ...m 2 Connect the load to the unit s RF OUTPUT connector 3 Set the REMOTE LOCAL switch to LOCAL 4 Check to see that the MAIN POWER switch on the unit s rear panel is set to the on position 5 Press the P...

Страница 28: ...tion screen Section 3 3 1 5 4 To get there from the Main Menu touch the User menu button followed by the I O menu button and finally the GPIB menu button Touching any of the buttons labeled 1 to 30 im...

Страница 29: ...TIMEOUT_ERROR message followed by a Line Feed is sent out from this port The baud rate for the Fiber Optic Serial port is user selectable from the RS 232 Fiber Optic Serial Baud Rate selection screen...

Страница 30: ...ated with a Line Feed character The cable required to make this connection is a USB 2 0 A B peripheral device cable The cable can be no longer than 5 meters If a longer distance is required a USB hub...

Страница 31: ...e Feed character is indicated by LF in subsequent command and query definitions All queries can be sent when the Toggle switch is in the LOCAL or REMOTE position All commands can only be sent when the...

Страница 32: ...Parameters State x OFF power off ON power on Response Format None No query for this command Example To turn the power on send the following command POWER ON LF To turn the power off send the followin...

Страница 33: ...nd the following command RESET LF 2 4 6 4 Level Adjust This command sets the RF gain of the 175A400 Syntax LEVEL xy Parameters Parameter x GAIN RF Gain Value y For RF Gain 0 Minimum 100 Maximum Respon...

Страница 34: ...ntity of the 175A400 send the following command IDN LF Response AR RF MICROWAVE INST XXXXXXXX 1 0 LF 2 4 6 6 IO Board Firmware Revision Query to get the firmware revision of the I O Board Syntax IOB P...

Страница 35: ...ions can be found in the table below NOTE Bits labeled NOT USED may be read as a bit state of 1 or 0 Table 2 5 States BIT STATE BIT POSITION BIT DESCRIPTION 0 1 NOTES x 0 NOT USED 1 NOT USED 2 NOT USE...

Страница 36: ...Gain 2 4 6 9 Faults Query to find the faults that have occurred with the 175A400 Syntax FSTA Parameters None Response Format FSTA space 00xx Where xx 00 to 14 Hexadecimal Table 2 6 Faults xx Decimal...

Страница 37: ...ing query OH LF Response OH space space space space 37 LF The system has spent 37 Hours in an RF On state 2 4 6 11 Operating Hours Power On Query to get the Power On operating hours Syntax OHP Paramet...

Страница 38: ...x firmware revision Example To get the firmware rev of the piggyback SBB assembly send the following command SBB LF Response SBB_SW_REV2 60 LF 2 4 6 13 System Serial Number Query to get the serial nu...

Страница 39: ...s back to their factory defaults 2 All applicable defaults can be queried except DEFAULT FACTORY by adding a character in place of the setting parameter 3 If the ALC Lockout Feature not available on a...

Страница 40: ...circuit from interlock connector pin 1 to pin 8 is required for normal operation Opening the Inhibit Interlock connection will inhibit the amplifier and display Fault Interlock on the front panel Touc...

Страница 41: ...U4 a resistive splitter R13 R14 R16 an RF power detector U6 and associated control circuitry The overall gain of the pre amplifier assembly is approximately 22 dB at minimum attenuation The power dete...

Страница 42: ...U1 and generates a fault signal if the drain current varies outside normal operating conditions It also monitors the module temperature via a thermistor R31 and generates a fault signal if the module...

Страница 43: ...onitor and indicate the status of the amplifier Power is supplied using only a single 5 volt power supply The board offers the following Feature Quantity Open drain outputs 4 Digital outputs 6 Digital...

Страница 44: ...Model 175A400 26 Rev C...

Страница 45: ...e to the importance of the amplifier s alignment it is recommended that when failure is caused by breakdown of any of the components in the signal circuits the amplifier be returned to the factory for...

Страница 46: ...described in the sections that follow 4 3 1 General Reading Faults 4 3 2 The Unit Cannot be Operated Remotely 4 3 3 Thermal Fault 4 3 4 Interlock Fault 4 3 5 Amplifier Faults 4 3 6 Low or No Power Ou...

Страница 47: ...24 25 26 or 27 should be 0 1V 3 If the voltage on A12 J5 pin 23 24 25 26 or 27 is high check the connection to J3 5 and C53 4 3 4 Interlock Fault Schematic Diagram No 10044790 The Model 175A400 is eq...

Страница 48: ...control is it set to maximum gain 2 Check the RF input to the unit is it the correct amplitude and frequency 3 Check the RF output connection from the unit is it correctly connected to the load Is the...

Страница 49: ...before conducting the RF tests specified in the following sections 1 The Model 175A400 s typical gain response at 20 dBm input and 0 dBm input is shown in Figure 4 1 and Figure 4 2 The actual gain may...

Страница 50: ...Model 175A400 32 Rev C Figure 4 3 Typical A Module Response 2 The typical response for the A1 Pre Amplifier at maximum gain setting is shown in Figure 4 4 Figure 4 4 Typical A1 Pre Amplifier Response...

Страница 51: ...cover If the security label is removed and or cut the warranty of the module will be voided Vacuum tubes in the L series amplifiers traveling wave tubes in TWT amplifiers and power heads will be free...

Страница 52: ......

Отзывы: