Dispay
Meaning
Action required
F 1
No measurement figure present
F 2
No „start-value” present, so
no actual BOD value can be
calculated
see „Autostart“
F120
An error has occurred on „Start“
Bottles must not be removed
during the „start“ process. Also,
positions must not be changed
during the „start“ process
Set DAte
During initial set up or after power
supply fails the time and date
must be reset
See „Setting Time and Date“
(parameters menu 3.11.1)
Srt und LoAd
The unit has not located a head
at the position which has been
selected
• Insert a bottle with head and press
the On/Off key; or :
• Select another position. The unit
will change to „Display Mode“
Srt und DEL
The data for the head at the
selected position have not yet
been read out.
• To continue the „start“ process,
press the „Enter“ key.
(NOTE :
All data on this head will be
deleted !)
• To quit the „start“ process, press
the On/Off key. The unit will then
change to „Display Mode“
End und LiFE
The electronics on the head
cannot store any more values
(this should not occur in normal
use).
The values most recently stored
can be read out. The head must
be replaced.