STAC2
- 50 -
9.7 CHANNELS CONFIGURATION PAGE
The channels configuration page display the main configuration related to physical channel’s
parameters measurement analysis and their hardware configuration. Each channel is materialized by a
sample suking thru a pinch valve and pump. Each pinch valve are multiplexing the differents fluid to
select correct fluid to suck thru the flowcell.
Customer is also able to select a list of analytic parameters he want to measure for each sample channel.
Enable channel #
Channel enabling
Must be done in
order. This it’s not
possible to enable
channel 1 & 3 for
example.
You must enable
channel :
1 or
1 & 2 or
1 & 2 & 3 or
1 & 2 & 3 & 4
Others combination
are not allowed
Set the frequency
measurement. This is
a multiple of main
cycle process
measurement
defined in general
configuration.
Ex:
10min main cycle
and frequency of 2
result in a sample
measurement every
20 min
Each channel are
independent.
Enable
dilution if
customer
know sample
concentration
level is too
high.
Sample
dilution will
be done with
Blank fluid.
Set the
dilution level:
0 = no
dilution
1 = 1 blank /
1 sample
…
Total tube length
for the related
channel in meter.
Pump suking
duration will be
automatically
adjusted
according tube
length outside
the STAC2 for
each channel.
Consequently,
sample
automatically
reach the flow
cell correctly.
Edit analytic
chemical
parameters
to measure.
Содержание Secomam STAC2
Страница 1: ...USER MANUAL version 1 03 ON LINE ANALYZER ...
Страница 2: ...STAC2 2 INSTALLATION USER S MANUAL ...
Страница 5: ...STAC2 5 ...
Страница 7: ...STAC2 7 ...
Страница 11: ...STAC2 11 ...
Страница 13: ...STAC2 13 ...
Страница 14: ......
Страница 27: ......
Страница 31: ......
Страница 59: ...STAC2 59 ...
Страница 65: ...STAC2 65 ...
Страница 74: ......