5.5.2 MODBUS RTU (RS485)
The RS485 connection allows the analyzer to communicate with MODBUS protocol on a RS485
bus.
5.5.2.1 Pins connection details
N° pin DB9 connector
Color
Designation
Signification
5
Black
GND
Ground
2
Orange
RS485_A
MODBUS A signal
3
Brown
RS485_B
MODBUS B signal
6
Red
12VDC / 0.5A(max)
Optional 12VDC output
.
Fig. 10 RTU MODBUS RS485 port pinout
5.5.3 DATA SIZE
1 START BIT - 8 S BITS - 1 STOP BIT - NO PARITY
Comply with MODBUS RTU protocol definition
5.5.4 DATA TRANSMISSION RATE
Bi-directional 9600 Baud.
IMPORTANT:
The distance between the STAC2 and the computer using the RS485 output must be lower than
30 meters.
Otherwise use on line amplifiers.
The standard RS485 is compatible with MODBUS protocol only !
.