Specification 

Power supply voltage 

9 to 26 volts, DC 

Supply current 

10mA average 

150mA peak (for 60 milliseconds during measurement) 

Output Signal 

0 to 5 volts logic signal 

Minimum load resistance 

5 k

 

Time to measure 

0.5 seconds after power is applied 

Time between measurements 

1 minute 

 

 

Sensor faults 

If a fault occurs, the sensor outputs a special signal of less than 10ppm, as follows: 

 

Frequency 

Fault condition 

0 ppm 

A broken wire or total failure of the sensor. Check the wiring and 
repair. Contact your distributor if wires are intact. 

2.5 ppm 

Moisture reading out of expected range. If this output continues, 
contact your distributor. 

5 ppm 

Low battery or supply voltage. Replace the batteries or check 
power supply. 

7.5 ppm 

Critical sensor settings have been lost. Contact your distributor. 

 

Conditions of Use 

Aquaflex must be installed as specified.  Use of Aquaflex data is entirely at the discretion of the 
user and should therefore be subject to current best practice principles of soil moisture 
management and agronomic management.  
 
Neither Streat Instruments nor its Distributors shall be liable (whether in contract, tort or otherwise) 
for any loss (including but not limited to loss of profits and consequential loss) of any kind whatever 
arising out of any published material or in connection with the performance or use of Aquaflex.  
 
The serial number marked on each Aquaflex Sensor must be recorded and quoted for warranty 
claims.  

 
 
 
 

Streat Instruments Limited

4A Expo Place
PO Box 24071

Christchurch

New Zealand

Envirofactors Limited
3 Water Lane
Bradford BD1 2JL
United Kingdom

Ph: +64 3 384 8900

Fax: +64 3 384 8901

[email protected]

www.streatsahead.com

Ph:  +44 1274 733 145
Fax: +44 1274 732 410

[email protected]
www.envirofactors.com

Отзывы: