Aquacal 75090 Скачать руководство пользователя страница 28

28

TROUBLESHOOTING

There are no warning
messages on the
display but the
chlorine level is too
low.
Water quality looks
dirty or cloudy.

Normal display

1. Check pool chemistry parameters. (See Water Balance & Chemistry

Recommendations in the Owner’s Quick Start section of manual.)
The cyanuric acid level may be low and the chlorine is being consumed
quickly by the UV from the sun.

2. The chlorine output needs to be increased.

There are no
warning messages
on the display.
The chlorine level is
too low but the pool
water looks fine.

1. The test kit reagents or strips may be old or have been exposed to

sunlight. Replace the kit or reagents and retest.

2. There is too much chlorine in the pool. The chlorine is bleaching the

test kit reagents. Dilute the water sample with distilled water and

retest.

Lower the chlorine output setting with

the down arrow button if the

chlorine level is

too high.
3. Sodium bromide may have been introduced into the pool by using a

bromine-based algaecide. The DPD (red color) chlorine test reagent
will give false readings if bromine is in the water. The OTO (yellow
color) test kit must be used which can test for bromine and chlorine.

Blank display

The Pool Pilot
Display is blank.

1. If the display is in bright sunlight, then shade the display to read.
2. Verify external time clock has not turned off  power to Control Unit.

(Temporarily override the time clock, if desired, to check the AG.)

3. Verify local shut off switch and/or main circuit breaker for Control

Unit is turned on.

4. If power is provided to  unit by an external control device, verify power

is provided to and from the device.

5. Fuse may be blown. See “Maintenance” for information on fuse replacement.
6. Check whether the GFCI breaker tripped and reset.

LED alternates red &
green flashing

Unit is not
generating Chlorine

Internal temperature of unit has exceeded shut off  temperature. Will
turn off chlorine generation for five (5) minutes or until temperature decreases.

Message Displayed

Problem

Typical Solution

1. Wait until unit cools down.
2. Move control unit to a shaded area if too hot.

Racetrack Pattern

Screen saver is
active

The screen saver activates if there are no warnings and no activity is
detected for > 30 minutes. Chlorine generation continues.

• Use the up arrow key to increase the chlorine output setting.

• Increase the pump run time so the chlorinator is generating chlorine

for a longer period of time.

3. Test water for high phosphate levels. Use a product such as LoChlor

Starver to reduce phosphates if the phosphate level is higher than
22-ppm (22 mg/l).

4. For Standard Salt Obtain a salt sample reading  and add salt, if

needed.

5

For E4aqua application obtain a salt sample reading with a test strip
and use a TDS meter to check the conductivity. If either is low adjust
either salt or additive as directed.

Содержание 75090

Страница 1: ...PMENT OWNER PN LTP0059 REV 1 03 10 2010 Owner s Manual Salt Chlorine Generator IMPORTANT Swimming Pool Spa Purification System Models 75090 75090 TL and 75091 Read This Manual Before Installing or Ope...

Страница 2: ...ell Manifold Assembly 75094 10 Mounting the Control Unit 11 Flow Rates 11 Electrical Connections 12 Grounding and Bonding 12 High Voltage Wiring 12 Twist Lock Connector 12 Three Prong Connector 12 Har...

Страница 3: ...r Chemistry E4aqua 21 Saturation Index SI E4aqua 22 MAINTENANCE 23 Fuse Location and Ratings 23 Testing the Flow Switch 23 Servicing the SuperCell 24 Removal of the SuperCell 24 Visual Inspection of t...

Страница 4: ...ours our voice mail system will handle your call Please be sure to leave your name a complete address and your telephone number You may also reach us by email at AutoPilotTechSupport teamhorner com Or...

Страница 5: ...e note WARNING relates to personal safety while CAUTION signals promote avoiding damage to equipment Safety Signals Safety Information Main Table A warning signal appears in this manual where special...

Страница 6: ...o heed the following may result in permanent injury or death WARNING To permit proper air circulation the Control Unit must be mounted at least 1 foot 300 mm off ground level Failure to heed the follo...

Страница 7: ...irculation pump running to operate For all other cells the water circulation pump must be operating for your AG to produce chlorine so run time is one of several key components to maintaining the prop...

Страница 8: ...ssembly 75094 The manifold is connected into the plumbing after all other equip ment Water from the pool spa is moved though the manifold by the circulation pump Replacement Parts RC11 Cell or RC 35 2...

Страница 9: ...et water level fall below level of Above Ground Manifold cell Damage to cell and power supply can occur if unit is run without water flow Failure to heed the following may result in equipment CAUTION...

Страница 10: ...aintenance The cables are 12 long The direction of the water flow through the manifold must be as indicated for the system to operate properly STEP 3 Install the 75093 into the return line hole and ti...

Страница 11: ...as applicable Using a screwdriver run the screws into the anchors leaving a gap between the wall and the underside of the screw heads 3 Holding the power center slotted openings to the screw heads al...

Страница 12: ...xterior of the control center The AG 75090 and 75090 TL have been configured for 115 VAC operation The AG 75091 has been factory configured for 230 VAC operation The configuration can be changed in th...

Страница 13: ...gently but firmly to connect BASE OF AG 3 The Flow Switch cable is used with the 75094 Inline cell manifold system This cable connector is also keyed and must be aligned to connect properly DO NOT us...

Страница 14: ...x 785 Length x Width x Average Depth x 785 Rectangular Round Oval Rectangular Round Oval Gallons pool size measured in feet Length x Width x Average Depth x 7 5 Diameter x Diameter x Average Depth x...

Страница 15: ...ould not be used as these mixtures contain high levels of impurities and may cause staining Note While not recommended granular salt containing anti caking additives such as YPS Yellow Prussiate of So...

Страница 16: ...olved into the pool 5 Use the UP and DOWN arrow buttons to set the purifier percentage to 50 and then allow the pool to operate normally Display will show P 50 scrolling once output is adjusted 6 For...

Страница 17: ...930 28 395 37 860 47 325 56 790 66 255 75 720 0 25 63 125 188 250 313 376 438 501 11 28 57 85 114 142 170 199 227 250 23 57 115 172 230 287 344 402 459 0 25 10 26 52 78 104 130 156 182 208 500 21 52 1...

Страница 18: ...dium bisulfate Low TA Eye irritation pH bounce stained etched plaster and metal corrosion Low TA Add sodium bicarbonate Total Alkalinity 80 to 100 ppm Monthly High TA Constant acid demand difficulty i...

Страница 19: ...um Hardness 400 ppm Total Alkalinity 125 ppm Salt Level 3000 ppm REFERENCE SECTION Use the chart below to determine your overall water balance Test water for pH water temperature Calcium Hardness Tota...

Страница 20: ...l then proceed to step 2 If current salt level is 750ppm or higher drain entire pool and refill with fresh water and proceed to step 2 2 With desired water level begin circulation pump Example An 11 0...

Страница 21: ...Ph High TA Add sodium bisulfate to correct for pH Do not add muriatic acid Do not add sodium bisulfate to a point where the pH drops below 7 2 Low Salt Below 3 mS leads to insufficient power to the ce...

Страница 22: ...of water that is whether your water will precipitate dissolve or be in equilibrium with calcium carbonate Water is properly balanced if the SI is 0 0 3 If SI is greater than 0 3 scaling and staining...

Страница 23: ...lease verify that the flow switch jumper is NOT installed then have the unit serviced 4 Turn on the water pump The red Check System light should turn off showing the unit has water flow 5 Turn off wat...

Страница 24: ...uld be straight and clear of any debris on the ends or between the blades END VIEW OF THE SUPERCELL LOOKING AT THE BLADES Clean the cell immediately and determine the cause of scaling See Basic Water...

Страница 25: ...vice CAUTION The electrical terminals must be completely dry to avoid corrosion and failure of the cell or cable Manual Cleaning of the SuperCell 1 Turn off circulator pump 2 Remove the cell and place...

Страница 26: ...ter and add the appropriate chemicals to balance the pool water per the levels recommended in the reference section on pages 17 22 Be sure to check salt and cyanuric acid stabilizer bringing those rea...

Страница 27: ...e Correct as appropriate If the volts are less than 20 then contact Autopilot Systems for assistance If this is a new installation Message Displayed Problem Typical Solution no FLO on 75094 Scrolling...

Страница 28: ...clock has not turned off power to Control Unit Temporarily override the time clock if desired to check the AG 3 Verify local shut off switch and or main circuit breaker for Control Unit is turned on...

Страница 29: ...receiver Connect the equipment into an outlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help FCC Compliance Decla...

Страница 30: ...30...

Страница 31: ...31...

Страница 32: ...32 2737 24th St North St Petersburg FL 33713 1 727 823 5642...

Отзывы: