HC997-999
– MOD. D
MAINTENANCE PROBE A - J
Digital multifunction solenoid metering pump
ENGLISH
ADSP9000033
03/09/2015 Rev.1.2
47/100
The membrane is extremely sensitive! When unscrew or screw the lid membrane ensure that the membrane not
trash with high or low pressure.
!
Unscrewing the membrane cap (11) without having exposed the valve opening (14), will destroy the
membrane!
Slightly lift the transparent covering (10) of the valve opening (14) close to the valve opening, using a small
screwdriver or similar tool, and push it a little to one side. This exposes the valve opening (14).
Unscrew the membrane cap (11) from the shaft (7) of the cell.
Now push the transparent covering (10) back into position, until it lies in the slot again and closes the valve
opening. Discard the used electrolyte. Clean the membrane cap (11) and G-holder (15
only for
ADELCJ15
) in clear water, then dry them.
Fill (11) and (15
only for ADELCJ15
) with the electrolyte supplied,
being careful not to create air
bubbles.
Only for ADELCJ15:
By putting the probe vertically, carefully push the measure(9) over the electrode cap
(15).
Screw the cap door (11) slowly and the cap (5) of the measuring cell, so that the electrolyte surplus can flow
from the valve open.
Screw the lid membrane completely. The O-Ring causes initial resistance. When the lid membrane is completely
screwed, cell measure should not beat against the membrane. This can damage and render it unusable.
( 15 only for ADELCLJ15)