The 400XAC includes

 TWO TEST MODES

 which change the way your AC Power Source operates.

 

 

 

 PROGRAM  MODE

 - Allows you to use test steps and memory locations within the AC Power 

Source menu. Designed for advanced testing applications.

 

 

 

 MANUAL  MODE

 - Allows you to run a test quickly using basic test parameters. Does not use 

preprogrammed test steps.

1.  Turn the power switch ON.  

2. From the Set screen, press the MORE soft key. In the next window press the SYSTEM soft key.

3.  In the System Parameters Menu, use the (

,

) arrow keys to highlight AUTO RUN parameter. 

Press EDIT to change the parameter.

4.  Press CHANGE to toggle the parameter between PROGRAM and MANUAL mode.

5. Press ENTER to save change.

6. Continue to adjust parameters according to testing needs. 

7. Press EXIT from System Parameters Menu to return to Set screen. 

When the AUTO RUN parameter is set to PROGRAM mode, the Set screen will be displayed as 
follows:

1. Turn ON the OC FOLD parameter

 
  From the Set screen, press the MORE soft key. In the next window press the SYSTEM soft key.
 
   In the System Parameters Menu, use the  (

,

)arrow keys to highlight the “OC FOLD” parameter. 

Press EDIT to change the parameter.

 
   Press CHANGE to set the highlighted parameter to ON. Press ENTER to save changes. Press EXIT 

twice to return to the Set screen.

2.  Program a current Hi-Limit, which represents the constant current you will provide to 

the load

 
  Press EDIT from the Set screen.

   Use the (

,

) arrow soft keys to highlight the “A Hi-Lmt” parameter. Press EDIT to change the 

parameter.

   Use the Numeric Keypad to enter the desired current limit. Press ENTER to save changes and 

then press EXIT to return to the Edit Parameters Menu. The limit set will be the constant current 
output.

3. Program the voltage

 
   From the Edit Parameters Menu, use the  (

,

) arrow soft keys to highlight the “Voltage” parameter. 

Press EDIT to change the parameter. 

   Using the Numeric Keypad enter the voltage which should be set higher than the nominal load 

voltage. This allows the current to adjust to the Hi-Limit value. 

   Press ENTER after you enter the voltage and press EXIT twice to return to the Set screen.

(System Parameters Menu)

(System Parameters Menu)

SELECT A TEST MODE

CONSTANT CURRENT MODE

WARNING:

 

LOCATE A SUITABLE TEST AREA WITH A THREE-PRONG, GROUNDED OUTLET. 

TEST YOUR THREE PRONG OUTLET FOR PROPER WIRING AND READ THE POWER 

CHECKLIST BEFORE USING YOUR INSTRUMENT.