How to adjust values via direct input:
1.
On the main screen, depending on the rotary
knob assignment, values can be set for voltage
(U), current (I), power (P) or resistance (R) via
direct input by tapping on the set/actual value
display areas, e.g. in the uppermost area of
voltage.
2.
Enter the required value using the decimal keypad. Similar to a pocket calculator,
the key
clears the input.
Decimal values are set by tapping the point key. For example, 54.3 V is set with
and
3.
The display then switches back to the main page and the set values take effect
6.4.7
Switching the main screen view
The main screen, also referred to as status page, with its set values, actual values and unit
status can be switched from the standard view mode with three or four values to a simpler
view mode with only voltage and current display. The advantage of the alternative view
mode is that actual values are displayed with much bigger characters, so they read be read
from a larger distance. Refer to section 6.4.3.9, “HMI / Front Panel Menu” on page 73 to see
where to switch the view mode in the MENU. Comparison:
Standard Display
Alternative Simpler Display
Limitations of the alternative display are:
Set points and actual values of power are not displayed and the set value of power
is only indirectly accessible
The set value of resistance is not displayed and only indirectly accessible
No access to the settings overview (MENU button) while the DC Input is on
Note: In alternative status page mode, the set values of power and resistance are not
adjustable while the DC Input is switched on. They can only be accessed and
adjusted in SETTINGS while the DC Input is off.