3A SERIES AC & DC LOAD OPERATION MANUAL
Page 46 of 56
7.4
Voltage Sense Input Terminals
To measure the UUT output voltage at the EUT terminals rather than the load input terminals,
the external voltage sense mode must be used. The external voltage sense wire harness is
provided for this purpose. It connects to the Voltage sense connector on the front panel of the
load module.
For small loads that deliver only low levels of current, INTERNAL sense can be used. For larger
loads that draw a considerable amount of current, EXTERNAL sense mode will compensate for
voltage drop along the load connection wires. Refer to Figure 7-1 below.
Figure 7-2: Internal and External Voltage Sense Connections