![Applied Photophysics Chirascan LD Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/applied-photophysics/chirascan-ld/chirascan-ld_user-manual_2955266012.webp)
Document 4207Q258 V3.00 January 2018
Page 12 of 30
Chirascan Couette Cell LD Accessory User Manual
2 PREPARATION
2.1
Removing the installed accessories
Remove any installed accessory, including all detectors. For instructions on removal of an accessory, see the
relevant user manual.
The Photomultiplier Tube (PMT) and Large Area Photodiode (LAAPD) detectors used with the
Chirascan spectrometers operate at high voltages and can produce a shock leading to serious
injury or death. Do not connect or disconnect the detector from the spectrometer unless the
spectrometer is powered off.
After the accessories have been removed, the interior of the spectrometer sample chamber should appear as
in Figure
Figure 2.1: The interior of the sample chamber after removal of accessories
2.2
LD calibration
The accessory needs to be calibrated before being used for linear dichroism measurements. If the accessory
is supplied with a new spectrometer, then the calibration will have been carried out by the Applied Photophysics
Production Department as part of the build and testing process. If it is supplied as an upgrade to an existing
instrument, then the calibration can be carried out by the user or by an Applied Photophysics engineer.
The principle of calibration is described by Norden [1]. An angled quartz fused silica plate is used as an
adjustable polarizing element. The LD standard used by APL is a 1 mm fused silica plate aligned at 30° to the
incident light. The ideal linear dichroism value for the 30° plate at 250 nm is
A = 0.03006.
Fluorescence port
locking ring