Image the Digital PCR 20K Chips

1.

Set the destination for the imaging data:

a.

From the Main Menu of the QuantStudio

 3D

Instrument, touch Destination.

b.

In the Destination screen, touch the desired destination

for the imaging data.

2.

If you choose to use a USB drive as a destination for your

imaging data, insert the drive into the USB port on the front

of the instrument.

3.

Open the chip tray and load the Digital PCR 20K Chip face-

up into the bay. Confirm that the Digital PCR 20K Chip is

correctly aligned within the chip tray, then close it.

X0300036

1

2

1

Orient the chip ID and fill port toward the front of the

instrument.

2

Confirm that the chip window is clean and correctly aligned

to the chip.

4.

In the instrument touchscreen, touch Start Run, then wait for

the QuantStudio

 3D Instrument to image the chip.

IMPORTANT! 

Do not open the chip tray or remove the USB

drive (if present) while the instrument displays the

countdown screen. Doing so will invalidate the image data

and require you to repeat the run.

5.

When the touchscreen displays the Analyzing Chip screen,

open the chip tray and remove the Digital PCR 20K Chip.

Store or discard it as desired.

You can read another Digital PCR 20K Chip without waiting

for the analysis to complete. To begin the next run, load the

chip and touch Start Run as described in step 3.

6.

After reviewing the results of the run, touch Done to close

the results screen.

If the analysis of the imaging data produced a   (red) for

either probe (FAM

 or VIC

®

 dye), visually inspect the chip

for problems and read the chip again.

After a Digital PCR 20K Chip has been read:

• You run the imaged Digital PCR 20K Chip again for up to

1 hour after thermal cycling.

• If you read multiple Digital PCR 20K Chips in rapid

succession, touch the scroll buttons ( / ) to review the

results of the previously imaged chips.

• The QuantStudio

 3D Instrument retains a copy of the

analyzed data for the imaged Digital PCR 20K Chip that you

can access from the Run History screen.

IMPORTANT! 

Confirm that the QuantStudio

 3D Instrument is

unloaded before you power off and move it. If the instrument is

moved while loaded, the Digital PCR 20K Chip inside can rotate

within the chip tray and obstruct the chip drawer, preventing it

from closing completely.

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

8

QuantStudio

 3D Digital PCR System Quick Reference

Содержание QuantStudio 3D

Страница 1: ...ital PCR reactions Required materials Genomic or complementary DNA sample s TaqMan Assay s for your experiment QuantStudio 3D Digital PCR Master Mix Gloves powder free nitrile Laboratory grade lint free wipes Microcentrifuge Permanent marker or pen Pipettes and tips P10 to P1000 Reaction tubes DNase RNase free non stick or low binding 0 5 mL or 1 5 mL TE Buffer 1X Molecular Biology Grade Vortexer ...

Страница 2: ...OpenArray Immersion Fluid Tip QuantStudio 3D Digital PCR Chip Loader UV Activated Chip Sealant Syringe Isopropanol Gloves powder free nitrile Scissors Laboratory grade lint free wipes Microcentrifuge Pipettes and tips P10 to P1000 Vortexer Load and seal the chips using a Chip Loader WARNING ULTRAVIOLET LIGHT HAZARD Looking directly at a UV light source can cause serious eye damage Never look direc...

Страница 3: ...g the chip into the chip nest as shown IMPORTANT Use the image within the chip nest to correctly orient the Digital PCR 20K Chip within the Chip Loader 2 1 c Press the sample loading blade lever then install the QuantStudio 3D Digital PCR Sample Loading Blade to the Chip Loader as shown IMPORTANT Press the Sample Loading Blade against the loader head to confirm that the blade is properly seated 2 ...

Страница 4: ...ll the port with Chip Sealant ensuring that the fluid touches the walls of the hole When full finish the seal by lifting the syringe tip and add a small amount of sealant over the top of the port X0300036 b Insert the Digital PCR 20K Chip assembly into the UV Curing Station on the Chip Loader Push the chip into the station until the ultraviolet light illuminates IMPORTANT Do not allow the Digital ...

Страница 5: ...of the assembled syringe then place it on a clean surface until ready for use IMPORTANT Use all of the Immersion Fluid within 60 minutes of uncapping the syringe Once a syringe is opened you cannot reattach the cap for later use 4 Remove the Chip Case Lids and Sample Loading Blades from their packaging and place them on a clean dry surface 5 Open the QuantStudio 3D Digital PCR 20K Chip package gen...

Страница 6: ...ng the syringe tip and add a small amount of sealant over the top of the port X0300036 b Power on the UV Stylus in the UV Stylus Stand by twisting the cap near the power button c Load the Digital PCR 20K Chip assembly into the UV Stylus Stand and allow the Chip Sealant to cure for 0 5 3 minutes A longer exposure no longer than 3 minutes ensures a complete seal X0300036 30 sec UV 10 Visually inspec...

Страница 7: ... Vol Stage 1 Stage 2 Stage 3 96 0 C 60 0 C 98 0 C 60 0 C 10 0 C Std 20 μL 10 00 2 00 0 30 2 00 99 59 1x 39x 1x You can remove the Digital PCR 20K Chips from the block immediately after the final extension step at 60 C is complete and the temperature of the block is 30 C IMPORTANT Read the Digital PCR 20K Chips within one hour after you remove them from the sample block After thermal cycling the Di...

Страница 8: ...nalyzing Chip screen open the chip tray and remove the Digital PCR 20K Chip Store or discard it as desired You can read another Digital PCR 20K Chip without waiting for the analysis to complete To begin the next run load the chip and touch Start Run as described in step 3 6 After reviewing the results of the run touch Done to close the results screen If the analysis of the imaging data produced a ...

Страница 9: ...ation for the QuantStudio 3D AnalysisSuite Software In order from high quality to low quality green Review of the analysis results is not required This flag is shown if the data meets all quality thresholds yellow Review of the analysis results is suggested This flag is shown if the Instrument cannot clearly identify the population of unamplified wells Distribution of unamplified wells on the chip...

Страница 10: ...nge and tips UV Curing Stylus for Chip Sealant UV Curing Stylus Stand 4488475 QuantStudio 3D Digital PCR UV Sealant Kit includes UV Activated Chip Sealant Syringe and tips 4486414 QuantStudio 3D Tilt Base for Dual Flat Block GeneAmp PCR System 9700 A25498 QuantStudio 3D Tilt Base for ProFlex 2x Flat PCR System For Research Use Only Not for use in diagnostic procedures The information in this guide...

Отзывы: