5-6 Creating and Editing Methods
Creating Methods
About the Default
Method
The GeneAmp PCR System 9700 comes with a default PCR thermal profile called a
method. The create screen displays this default method. For information on displaying
the create screen, see “Displaying the Create Methods Screen” on page 5-7.
You can run the default method shown above, or use it as a template to create a
customized method.
Basic Parameters
To create a method, you need to define the following four basic parameters:
♦
Temperature Control Parameters
♦
Pre-PCR Holds
♦
PCR Parameters
♦
Post-PCR Holds
For detailed information about these parameters refer to the table below:
Pre-PCR segment
PCR segment
Post-PCR segment
Number of
temperature
control
parameters
in the
pre-PCR
segment
Number of
temperature control
parameters in the
PCR segment
Number of cycles the
PCR segment
will run
Number of
temperature
control parameters
in the post-PCR
segment
1 Hld
3 Tmp
25 Cycles 2 Holds
94.0
94.0
5:00
0:30
55.0
0:30
72.0
0:30
72.0
7:00
4.0
F1
F3
F4
F5
F2
Start
Cancel
More
Store
For information about...
See page
Содержание GeneAmp PCR System 9700
Страница 1: ...GeneAmp PCR System 9700 Base Module User s Manual ...
Страница 9: ...ix To Obtain Customer Training Information C 7 D Limited Warranty Statement Index ...
Страница 10: ......
Страница 28: ......
Страница 56: ......
Страница 90: ......
Страница 96: ......
Страница 110: ......
Страница 116: ......
Страница 124: ......
Страница 126: ......
Страница 131: ......