3-10 Instrument Setup
Turning the
Run-Time Beeper
On or Off
When turned on, the run-time beeper beeps during a pause and once at the
completion of a run. The default value is Off.
Setting the Pause
Time Out
The Pause Time Out field sets the time in minutes:seconds format for the length of a
pause when you use the Pause soft key to pause a run from the Run Time screen. For
more information on Pausing a Run, see “Pausing a Run” on page 4-8.
To turn the Run Time-Beeper on or off:
Step
Action
1
Select the
Run Time Beep
field. This changes the value of the soft keys.
2
You can take the following action:
3
Press
Enter
to accept your entry.
If you want to...
Then...
turn the beeper on
Press ON.
turn the beeper off
Press Off.
To set the pause time out:
Step
Action
1
Select
More
to display the second configuration screen.
2
Set the
Pause Time Out
field by using the numeric keys to enter a time in minutes
and seconds.
From this screen, you can also define the:
♦
Idle state setpoint temperature.
♦
Baud rate for your printer port.
3
Press
Accept
when all information on this screen is correct.
Note
CE
clears an entry.
Instrument Configuration
Pause Time Out: 10:00 (00:01-99:59)
Idle State Setpoint: 25.0°C (4.0-99.9)
Baud Rate: 9600
F1
F3
F4
F5
F2
Accept
Cancel
More
Содержание GeneAmp PCR System 9700
Страница 1: ...GeneAmp PCR System 9700 Base Module User s Manual ...
Страница 9: ...ix To Obtain Customer Training Information C 7 D Limited Warranty Statement Index ...
Страница 10: ......
Страница 28: ......
Страница 56: ......
Страница 90: ......
Страница 96: ......
Страница 110: ......
Страница 116: ......
Страница 124: ......
Страница 126: ......
Страница 131: ......