Chapter 7 Data Explorer Examples
7-4
Applied Biosystems
7
7.1.2 Deconvoluting and
Evaluating Unresolved
Chromatographic Peaks
Overview
You can use the Data Explorer software to deconvolute
chromatographic peaks and obtain masses for each
component by:
•
Creating a combined spectrum to identify the coeluting
species by mass
•
Creating extracted ion chromatograms for the masses,
and comparing the extracted ion chromatograms to the
original TIC containing the unresolved peaks
You can then create a combined spectrum from each
extracted ion chromatogram to evaluate the spectral data.
Creating a
combined
spectrum
The following example illustrates how to deconvolute two
unresolved chromatographic peaks in a tryptic digest of
cytochrome c:
1.
Display the data file.
2.
In the Chromatogram window, zoom in on the unresolved
peaks (Figure 7-2).
Figure 7-2 Unresolved Peaks in Cytochrome C
3.
In the Chromatogram window, right-click-drag over the
unresolved peak pair.
Содержание Data Explorer 4 Series
Страница 1: ...Data Explorer Software Version 4 Series Software User Guide ...
Страница 10: ...Table of Contents x Applied Biosystems ...
Страница 56: ...Chapter 1 Data Explorer Basics 1 42 Applied Biosystems 1 ...
Страница 98: ...Chapter 2 Using Chromatogram and Spectrum Windows 2 42 Applied Biosystems 2 ...
Страница 172: ...Chapter 3 Peak Detection and Labeling 3 74 Applied Biosystems 3 ...
Страница 204: ...Chapter 4 Examining Chromatogram Data 4 32 Applied Biosystems 4 ...
Страница 270: ...Chapter 5 Examining Spectrum Data 5 66 Applied Biosystems 5 ...
Страница 316: ...Chapter 6 Using Tools and Applications 6 46 Applied Biosystems 6 ...
Страница 384: ...Chapter 9 Troubleshooting 9 22 Applied Biosystems 9 ...
Страница 388: ...Appendix A Warranty A 4 Applied Biosystems A ...
Страница 416: ...Appendix C Data Explorer Toolbox Visual Basic Macros C 20 Applied Biosystems C ...
Страница 446: ...I D E X N Index 30 Applied Biosystems ...