![Applied Biosystems Data Explorer 4 Series Скачать руководство пользователя страница 180](http://html.mh-extra.com/html/applied-biosystems/data-explorer-4-series/data-explorer-4-series_user-manual_2955117180.webp)
Chapter 4 Examining Chromatogram Data
4-8
Applied Biosystems
4
From the
Spectrum window
To create an extracted ion chromatogram for a mass range
from the Spectrum window:
1.
Click the Chromatogram window to activate it.
2.
Select Duplicate Active Trace from the Display menu to
keep the original data displayed after processing.
3.
In the Spectrum window, right-click-drag over the mass
region of interest in the extracted ion chromatogram. The
width of the box you draw defines the precise mass range
used in the extracted ion chromatogram.
The extracted ion chromatogram is displayed in the
Chromatogram window (Figure 4-2) with the mass
range indicated in the trace label.
Figure 4-2 Extracted Ion Chromatogram
4.
To return to the original trace, see “Returning to the
original trace” on page 4-4.
Extracted ion chromatogram
(center window)
Содержание Data Explorer 4 Series
Страница 1: ...Data Explorer Software Version 4 Series Software User Guide ...
Страница 10: ...Table of Contents x Applied Biosystems ...
Страница 56: ...Chapter 1 Data Explorer Basics 1 42 Applied Biosystems 1 ...
Страница 98: ...Chapter 2 Using Chromatogram and Spectrum Windows 2 42 Applied Biosystems 2 ...
Страница 172: ...Chapter 3 Peak Detection and Labeling 3 74 Applied Biosystems 3 ...
Страница 204: ...Chapter 4 Examining Chromatogram Data 4 32 Applied Biosystems 4 ...
Страница 270: ...Chapter 5 Examining Spectrum Data 5 66 Applied Biosystems 5 ...
Страница 316: ...Chapter 6 Using Tools and Applications 6 46 Applied Biosystems 6 ...
Страница 384: ...Chapter 9 Troubleshooting 9 22 Applied Biosystems 9 ...
Страница 388: ...Appendix A Warranty A 4 Applied Biosystems A ...
Страница 416: ...Appendix C Data Explorer Toolbox Visual Basic Macros C 20 Applied Biosystems C ...
Страница 446: ...I D E X N Index 30 Applied Biosystems ...