Chapter 3 Peak Detection and Labeling
3-28
Applied Biosystems
3
Advanced
Settings
(spectrum data
only)
Table 3-4 describes the parameters in the Advanced Settings
tab of the Spectrum Peak Detection Setup dialog box (see
Figure 3-6 on page 3-17).
Default peak detection values are listed in Section 3.7, Default
Peak Detection Settings.
NOTE: If you select Use Resolution Dependent Settings in
the Basic Settings tab, Basic Settings override Advanced
Settings. The Advanced Settings tab is accessible, but all
parameters on the tab are dimmed. To make Advanced
Settings available for editing, select Use Advanced
Settings on the Basic Settings tab.
Table 3-4 Advanced Settings (Spectrum Data Only)
Parameter
Description
Peak Detection Settings
Detection Ranges
Specifies the region of the trace (defined by x-axis lower and
upper boundaries) to which the settings apply.
If you select Use Resolution Dependent Settings in the Basic
Settings tab (described on page 3-23), the software
automatically divides the trace into detection ranges. If you
select Use Advanced Settings, the detection ranges
calculated by the software are maintained, and you can
modify the ranges and settings, or combine all ranges in the
list into one range, as described below.
Continued
Содержание Data Explorer 4 Series
Страница 1: ...Data Explorer Software Version 4 Series Software User Guide ...
Страница 10: ...Table of Contents x Applied Biosystems ...
Страница 56: ...Chapter 1 Data Explorer Basics 1 42 Applied Biosystems 1 ...
Страница 98: ...Chapter 2 Using Chromatogram and Spectrum Windows 2 42 Applied Biosystems 2 ...
Страница 172: ...Chapter 3 Peak Detection and Labeling 3 74 Applied Biosystems 3 ...
Страница 204: ...Chapter 4 Examining Chromatogram Data 4 32 Applied Biosystems 4 ...
Страница 270: ...Chapter 5 Examining Spectrum Data 5 66 Applied Biosystems 5 ...
Страница 316: ...Chapter 6 Using Tools and Applications 6 46 Applied Biosystems 6 ...
Страница 384: ...Chapter 9 Troubleshooting 9 22 Applied Biosystems 9 ...
Страница 388: ...Appendix A Warranty A 4 Applied Biosystems A ...
Страница 416: ...Appendix C Data Explorer Toolbox Visual Basic Macros C 20 Applied Biosystems C ...
Страница 446: ...I D E X N Index 30 Applied Biosystems ...