1.
Click Refresh to update consumable status.
The Consumables pane displays expiration dates and lot numbers (determined
from the RFID tags on the consumable containers).
Note:
The Expiration Date for consumables is displayed in red if the
consumable is within the following days of expiration: Pouch 7 days, Buffers
7 days, Capillary array 1 day.
2.
Check the consumables gauges for the number of injections, samples, or days
remaining for a consumable. See “Instrument consumables handling, usage
limits, and expiration“ on page 19 for information.
When <10% of the allowed use of the consumable remains, the gauge moves into
the red warning range. The consumable also displays in red in the Consumables
pane.
IMPORTANT!
We recommend that you add a calendar reminder to the schedule
for polymer and buffer replacement. Set the notification to display 2 days before
the polymer should be replaced.
Check
consumables
status
Chapter 2
Start the system
Check system status in the Dashboard
2
3500/3500xL Genetic Analyzer User Guide—Data Collection Software v3.1
33